Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1008-2025

Brand Name: N/A Product Name: Repeater Bone Tamp Recall

Brand Name: N/A Product Name: Repeater Bone Tamp Model/Catalog Number: ASY-00132 Software Version: N/A Product Description: The Repeater Bone Tamp is a cylindrical shaft used to push bone graft material through a corresponding bone funnel into the FlareHawk interbody fusion device in spinal fusion procedures. Component: No

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: N/A Product Name: Repeater Bone Tamp. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1008-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 20, 2024
FDA report date
Jan 29, 2025
Quantity in commerce
9
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Brand Name: N/A Product Name: Repeater Bone Tamp Model/Catalog Number: ASY-00132 Software Version: N/A Product Description: The Repeater Bone Tamp is a cylindrical shaft used to push bone graft material through a corresponding bone funnel into the FlareHawk interbody fusion device in spinal fusion procedures. Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-1008-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
ASY-00132Brand Name: N/A Product Name: Repeater Bone TampCheck source noticeNot specified
Not specifiedBrand Name: N/A Product Name: Repeater Bone TampCheck source notice00810004721906
FDA code information

Lot Code: Part Number: ASY-00132 Lot Number: 03200045 UDI Number: (01)00810004721906(10)03200045

View the original FDA recall notice

Why was this product recalled?

Due to incompatibility of bone tamp (Rev A) and bone funnels (Rev C).

What actions did Integrity Implants Inc. take?

On 12/20/2024, the firm emailed a "NOTICE OF RECALL" letter to customer informing them that the firm has received a report relating to a specific lot of Repeater Bone Tamps not fitting with newer lots of the Repeater Bone Funnel. Customers are instructed to: 1. Segregate Product. immediately remove all Repeater Bone Tamp (Part # ASY-00132, Lot Number 03200045), from their inventory (regardless of its location) and segregate this product in a secure location for return to Accelus. 2. Complete QF-00070 Product Tracking & Verification Form. Complete and return the enclosed QF-00070 Product Tracking & Verification Form (even if you do not have any product to return), following the directions on this page and QF-00070 Product Tracking & Verification Form. Accelus' customer service representative can assist in completing the form. 3. Indicate Returned Products on QF-00070 Product Tracking & Verification Form. QF-00070 Product Tracking & Verification Form lists the Repeater Bone Tamp and product descriptions. Mark the number of each type of Repeater Bone Tamp that will be returned on this form. 4. Call (561)529-3861 or 1-800-201-9300 from 8:30am to 5pm (EDT) should there be any questions about returning the product. Customer's Accelus account will be credited in an amount equal to the actual purchase price for the returned product, if applicable. Inform Customer Service Representative whether customers would like to order a replacement product. 5. Package the Return Material. Package Repeater Bone Tamp(s) that are being returned in an appropriate shipping box. Place a signed original copy of QF-00070 Product Tracking & Verification Form in the box, appropriately seal the box and send it on or before 12-26-2024, to: Accelus 354 Hiatt Drive, Suite 100 Palm Beach Gardens, FL 33418 Office: (561)529-3861 or 1-800-201-9300 Fax: (561) 529-3869

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology