Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1017-2025

Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Recall

Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP3403

FDA recall Class II

Ongoing

FDA lists this Class II recall for Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1017-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 20, 2024
FDA report date
Jan 29, 2025
Quantity in commerce
15 kits
Distribution
See key details
FDA cause
No Marketing Application

What is being recalled?

Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers. Catalog Number: LP3403

Affected model & UDI checker

Search the exact identifiers published for Z-1017-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
LP3403Randox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers.Check source noticeNot specified
Not specifiedRandox Lipoprotein (a) - In vitro diagnostic (IVD) use as Randox Lipoprotein (a) assay on automated analysers.Check source notice05055273204421
FDA code information

UDI-DI: 05055273204421 All lots within expiry

View the original FDA recall notice

Why was this product recalled?

Lipoprotein (a) on Roche Cobas c501 with nmol/l claims does not have a 510(k) or PMA.

What actions did RANDOX LABORATORIES LIMITED take?

Randox Laboratories Ltd (manufacturer) extended recall to the distribution center in the USA via email on 12/20/24. The distributor will then contact the customer directly via email. Letter states reason for recall, health risk and action to take: Action to be taken: " Discontinue use of and discard any copies of Roche Cobas c501 ISA listing nmol/1 claims. " Ensure clinical results are no longer reported in nmol/1 using the Roche Cobas c501 nmol/1 claims. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organisation. If you have any questions or concerns please contact Randox Technical Services.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.