Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration. Recall

Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

FDA recall Class I

Ongoing

FDA lists this Class I recall for Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1019-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jan 10, 2025
FDA report date
Feb 12, 2025
Quantity in commerce
23 systems
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Affected model & UDI checker

Search the exact identifiers published for Z-1019-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedIvenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.Check source notice00811505030122
FDA code information

UDI-DI: 00811505030122. Software 5.9.2 and prior

View the original FDA recall notice

Why was this product recalled?

Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Fresenius Kabi USA, LLC take?

Fresenius Kabi began notifying consignees on 01/10/2024 via emailed letter. Consignees were instructed to check if your device contains the affected software, review and become familiar with the temporary Risk Mitigation actions for each anomaly outlined in the letter and notify customers/personnel if the product was transferred or further distributed. Additionally, to facilitate the installation of the pump software, LVP SW 5.10, your facility must install the new Ivenix Infusion Management System (IMS) software version (5.2) on your IMS server and install the new software version (5.10) for the Ivenix Large Volume Pump (LVP). Please post this letter in areas that affected devices are stored and used such that users of this device are aware of the issues and recommended risk mitigation measures. Lastly, consignees were requested to complete and return the Customer Reply Form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology