Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. Recall
Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. Model/Catalog Number: 530159
FDA recall Class II
Ongoing
FDA lists this Class II recall for Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1021-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 4, 2024
- FDA report date
- Feb 5, 2025
- Quantity in commerce
- 1090 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. Model/Catalog Number: 530159
Affected model & UDI checker
Search the exact identifiers published for Z-1021-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 530159 | Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. | Check source notice | Not specified |
| Not specified | Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal space and epidural space for pain control. | Check source notice | 04046964669286 |
FDA code information
UDI-DI 04046964669286 (Primary); 04046964669279 (Unit of Use); Lot Number 0061916572
Why was this product recalled?
Incorrect catheter connector within the kit resulting in inability to provide a secure connection to the epidural catheter. The incorrect catheter connector is smaller than the required catheter connector and therefore is unable to be used to administer medication.
What actions did B. BRAUN MEDICAL INC. take?
B Braun issued Urgent Medical Device Correction Notification letter on 12/04/24 via USPS Certified Mail. Letter states reason for recall, health risk and action to take: 1. Review the Medical Device Correction Notification in its entirety and ensure that all users in your organization of the above-mentioned product, and other concerned persons are informed about this voluntary correction. If you have further distributed the affected product, please forward this notice to your consignees. The correction is to be extended to the hospital/healthcare facility level. 2. Post this notification where the affected products are stored. Additional catheter connectors are being supplied to you and should be stored where the affected inventory is utilized. In the event that an incorrect catheter connector is identified, please utilize these additional devices to facilitate replacement. 3. Returned the completed Urgent Medical Device Correction Acknowledgement Form to B. Braun Medical Inc. Quality Assurance department by faxing the form to (610) 849-1197 or email to recalls@bbraunusa.com within two (2) weeks of receipt, even if the total inventory in your possession is zero (0). Adverse reactions or quality problems experienced with this product, or questions about this recall may be reported to BBMI s Postmarket Surveillance Department by calling 1-844-903-6417.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1021-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology