Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1024-2026

MAVERICK Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496 Recall

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

FDA recall Class II

Ongoing

FDA lists this Class II recall for MAVERICK Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1024-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 10, 2025
FDA report date
Jan 21, 2026
Quantity in commerce
33
Distribution
See key details
FDA cause
Radiation Control for Health and Safety Act

What is being recalled?

MAVERICK Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

View the original FDA recall notice

Why was this product recalled?

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did REPLIGEN CORPORATION take?

Repligen will repair the connections in each potentially affected MAVERICK device. In the meantime, all customers have been instructed to refrain from using the device.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.