Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1028-2026

BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03, 139676-01, 139421-01, 139420-01, 139419-01, 139419-02, 139673-01, 137081-01, 136451-01, 136451-02, 136450-01, 136450-02, 136449-01, 136449-02, 134148-01, 133760-01, 134149-01, 137669-02, 137669-03, 139675-01, 139672-01 139418-01, 139418-02, 136448-01, 136448-02, 137673-02, 137673-03 139422-01, 133744-01, 136452-01, 136452-02, 138551-01, 134157-01 Software Versions: System Versions 1.6.1 - 1.11.0 Server Application Versions ES 5.4X - 5.8.X The BD Pyxis Enterprise Server is a web application intended to configure and manage BD Pyxis ES automated dispensing cabinets (ADC) and associated BD PyxisTM ES products. Recall

BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03, 139676-01, 139421-01, 139420-01, 139419-01, 139419-02, 139673-01, 137081-01, 136451-01, 136451-02, 136450-01, 136450-02, 136449-01, 136449-02, 134148-01, 133760-01, 134149-01, 137669-02, 137669-03, 139675-01, 139672-01 139418-01, 139418-02, 136448-01, 136448-02, 137673-02, 137673-03 139422-01, 133744-01, 136452-01, 136452-02, 138551-01, 134157-01 Software Versions: System Versions 1.6.1 - 1.11.0 Server Application Versions ES 5.4X - 5.8.X The BD Pyxis Enterprise Server is a web application intended to configure and manage BD Pyxis ES automated dispensing cabinets (ADC) and associated BD PyxisTM ES products.

FDA recall Class II

Ongoing

FDA lists this Class II recall for BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03, 139676-01, 139421-01, 139420-01, 139419-01, 139419-02, 139673-01, 137081-01, 136451-01, 136451-02, 136450-01, 136450-02, 136449-01, 136449-02, 134148-01, 133760-01, 134149-01, 137669-02, 137669-03, 139675-01, 139672-01 139418-01, 139418-02, 136448-01, 136448-02, 137673-02, 137673-03 139422-01, 133744-01, 136452-01, 136452-02, 138551-01, 134157-01 Software Versions: System Versions 1.6.1 - 1.11.0 Server Application Versions ES 5.4X - 5.8.X The BD Pyxis Enterprise Server is a web application intended to configure and manage BD Pyxis ES automated dispensing cabinets (ADC) and associated BD PyxisTM ES products.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1028-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 3, 2025
FDA report date
Jan 14, 2026
Quantity in commerce
2,079 systems
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03, 139676-01, 139421-01, 139420-01, 139419-01, 139419-02, 139673-01, 137081-01, 136451-01, 136451-02, 136450-01, 136450-02, 136449-01, 136449-02, 134148-01, 133760-01, 134149-01, 137669-02, 137669-03, 139675-01, 139672-01 139418-01, 139418-02, 136448-01, 136448-02, 137673-02, 137673-03 139422-01, 133744-01, 136452-01, 136452-02, 138551-01, 134157-01 Software Versions: System Versions 1.6.1 - 1.11.0 Server Application Versions ES 5.4X - 5.8.X The BD Pyxis Enterprise Server is a web application intended to configure and manage BD Pyxis ES automated dispensing cabinets (ADC) and associated BD PyxisTM ES products.

Affected model & UDI checker

Search the exact identifiers published for Z-1028-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1115-00BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03, 139676-01, 139421-01, 139420-01, 139419-01, 139419-02, 139673-01, 137081-01, 136451-01, 136451-02, 136450-01, 136450-02, 136449-01, 136449-02, 134148-01, 133760-01, 134149-01, 137669-02, 137669-03, 139675-01, 139672-01 139418-01, 139418-02, 136448-01, 136448-02, 137673-02, 137673-03 139422-01, 133744-01, 136452-01, 136452-02, 138551-01, 134157-01 Software Versions: System Versions 1.6.1 - 1.11.0 Server Application Versions ES 5.4X - 5.8.X The BD Pyxis Enterprise Server is a web application intended to configure and manage BD Pyxis ES automated dispensing cabinets (ADC) and associated BD PyxisTM ES products.Check source noticeNot specified
FDA code information

Catalog Number: 1115-00 Software Versions: UDI-DI code for Hardware: 10885403518348 UDI-DI code for Software: 10885403520341 System Versions 1.6.1 - 1.11.0 Server Application Versions ES 4.13 - 5.8.X

View the original FDA recall notice

Why was this product recalled?

Due a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did CareFusion 303, Inc. take?

On 12/03/2025, the firm sent U.S. Postal and email an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them that BD received customer complaints of BD Pyxis" equipment not receiving timely updates of patient, medication, or related information from enterprise systems. This resulted in new patient and/or order information to be delayed from displaying on connected Pyxis equipment, and a delay in any changes to current patient orders being applied to the patient profile within Pyxis. Customers are instructed to: 1.Refer to the BD Pyxis" ES User Guides warning: Instruct end users to utilize the BD Pyxis" ES System patient/order information in conjunction with the medical record when making clinical decisions. Making clinical decisions solely from BD Pyxis" ES System information may result in errors in care resulting in serious injury or death. 2.BD Pyxis" MedStation" devices can be utilized in both profile mode (specific medications orders are associated with specific patients) and non-profiled mode. 3.Facilities should establish comprehensive policies and procedures for situations when an automated dispensing system is not receiving updated information and medications may not be loaded in device inventory. 4.Device keys can be used to manually access unavailable drawers. Keys must be readily available and retrievable to constitute risk mitigation. 5.Devices can also be placed into critical override mode, where all contents of the device will be available to users. (For additional detailed instruction - refer to Urgent Medical Device Correction letter. BD recommends that customers who are preparing to implement or upgrade to ES 1.11 continue as planned in order to receive the bug fixes it contains and prepare to receive the ES 1.11.1 patch when it is available. For Questions: BD Remediation Support: Email: US-Pyxis-Field-Action-Remediation@bd.com BD Technical Support: Website: bd.com/self-service, Phone: 1-800-727-6102, Hours: 24 hours, 7 days a week

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology