Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1029-2026

The Pressio 2 ICP Monitoring System, Model Number PSO-4000; Intracranial Pressure Monitor Recall

The Pressio 2 ICP Monitoring System, Model Number PSO-4000; Intracranial Pressure Monitor

FDA recall Class II

Ongoing

FDA lists this Class II recall for The Pressio 2 ICP Monitoring System, Model Number PSO-4000; Intracranial Pressure Monitor. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1029-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Nov 17, 2025
FDA report date
Jan 14, 2026
Quantity in commerce
105 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

The Pressio 2 ICP Monitoring System, Model Number PSO-4000; Intracranial Pressure Monitor

Affected model & UDI checker

Search the exact identifiers published for Z-1029-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
PSO-4000The Pressio 2 ICP Monitoring System, Model Number PSO-4000; Intracranial Pressure MonitorCheck source noticeNot specified
Not specifiedThe Pressio 2 ICP Monitoring System, Model Number PSO-4000; Intracranial Pressure MonitorCheck source notice03760124132076
FDA code information

UDI/DI 03760124132076, All monitors

View the original FDA recall notice

Why was this product recalled?

Customer complaints of Pressio monitor rebooting.

What actions did Sophysa take?

Sophysa issued a Field Safety notice to its consignees on 11/25/2025 via email. The notice explained the issue and requested monitoring of the patient should the unintended reboot of the monitor occur. The consignees were directed to notify their sales representatives to schedule the monitor update.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology