HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553540 Recall
HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553540
FDA recall Class II
Ongoing
FDA lists this Class II recall for HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553540. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1034-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 20, 2024
- FDA report date
- Feb 5, 2025
- Quantity in commerce
- 519 units (OUS)
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553540
Affected model & UDI checker
Search the exact identifiers published for Z-1034-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553540 | Check source notice | 08714729904564 |
FDA code information
GTIN: 08714729904564 Lot Numbers: 33169261, 33169263, 33209127, 33243075, 33243077, 33243856, 33255964, 33488328, 33787031, 33787032, 33853395, 33882221, 33882222, 33899656, 33909280, 33979053, 33979055, 34027245, 34027246, 34035523, 34035524, 34052776, 34052778, 34161831, 34161832, 34161833, 34161834, 34190943, 34190944, 34190945, 34192326, 34192327, 34201917, 34203537, 34203539, 34204795, 34223677, 34223679, 34226399, 34226540, 34243427, 34243428, 34375175, 34381803, 34585408, 34731098
Why was this product recalled?
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
What actions did BOSTON SCIENTIFIC CORPORATION take?
Boston Scientific issued Urgent Medical Device Removal - Immediate Action Required Letter via Fedex Priority Overnight on December 20, 2024 to US consignees. Letter states reason for recall, health risk and action to take: This removal affects only the products listed in Attachment 1. Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. Share this notice with any health care professional within your organization and with any organization where the affected products may have been transferred. " If you are a distributor, this notification must be forwarded to your customers to ensure notification of this product removal is carried out to the end-user level. If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. Contact your local Boston Scientific representative with any questions.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1034-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026