NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10 Recall

NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10

FDA recall Class II

Ongoing

FDA lists this Class II recall for NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1039-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 6, 2025
FDA report date
Feb 5, 2025
Quantity in commerce
155 units
Distribution
See key details
FDA cause
Process change control

What is being recalled?

NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10

Affected model & UDI checker

Search the exact identifiers published for Z-1039-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedNexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10Check source notice00889024635746
FDA code information

UDI-DI: 00889024635746; Lot Numbers:66782843 66782840 66873137

View the original FDA recall notice

Why was this product recalled?

It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.

What actions did Zimmer, Inc. take?

An URGENT MEDICAL DEVICE RECALL - REMOVAL notification letter dated 1/6/25 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to orporateQuality.PostMarket@zimmerbiomet.com. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products available to return in your territory. b. For International Returns, request an IRA by emailing zimmerbiometintlirareques

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.