A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung tablets (CFN CT900). Recall
A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung tablets (CFN CT900).
FDA recall Class II
Ongoing
FDA lists this Class II recall for A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung tablets (CFN CT900).. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1043-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 3, 2025
- FDA report date
- Jan 21, 2026
- Quantity in commerce
- 7123
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
A71200 Vanta" Clinician Programmer Application (CP App) version 2.0.2465 and 2.0.2683, Product number A71200. Software is deployed and operated on Samsung tablets (CFN CT900).
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
version 2.0.2465 and 2.0.2683 UDI-DI (GTIN) 00763000520083 Affected product is the software. Implanted Vanta Neurostimulator devices are not affected.
Why was this product recalled?
Complaints received that Vanta A71200 CP App does not function as intended during use. There is a potential for prolonged or postponed surgical procedure.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Medtronic Neuromodulation take?
On December 3, 2025 the firm distributed Urgent Medical Device Safety Notification letters to affected consignees. Customers were instructed to download and install A81200 CP App v2.0.2684 or any later version.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1043-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data