Infinity Central Station. Recall
Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Infinity Central Station.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1066-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 14, 2025
- FDA report date
- Feb 5, 2025
- Quantity in commerce
- 1,013 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Infinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.
Affected model & UDI checker
Search the exact identifiers published for Z-1066-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| MS26800 | Infinity Central Station. | Check source notice | Not specified |
| MS34036 | Infinity Central Station. | Check source notice | Not specified |
| Not specified | Infinity Central Station. | Check source notice | 04049098001878 |
| Not specified | Infinity Central Station. | Check source notice | 04048675548027 |
FDA code information
Infinity CentralStation (ICS) Model Number: MS26800. UDI-DI Number: 04049098001878. Edifier Speaker Model Number: MS34036. UDI-DI Number: 04048675548027. Serial Numbers: 12946M00269, 12946M00314, 12946M00506, 12946M00519, 12946M00518, 12946M00553, 1OQ7WR-00103, 1OQ8CA-00249, 1OQ8CA-00254, 1OQ8CA-00229, 1OQ8CA-00221, 1OQ8CA-00282, 1OQ8CA-00314, 1OQ8CA-00193, 1OQ8CA-00200, 1OQ8CA-00194, 1OQ8CA-00181, 1OQ8CA-00178, 1OQ8CA-00248, 1OQ8CA-00125, 1OQ8CA-00071, 1OQ8CA-00032, 1OQ8CA-00171, 1OQ8CA-00164, 1OQ8CA-00149, 1OQ8CA-00069, 1OQ8CA-00057, 1OQ8CA-00153, 1OQ8CA-00150, 1OQ8CA-00148, 1OQ8CA-00165, 1OQ8CA-00155, 1OQ8CA-00097, 1OQ8CA-00113, 1OQ8CA-00106, 1OQ8CA-00075, 1OQ8CA-00052, 1OQ8CA-00067, 1OQ8CA-00128, 1OQ8CA-00132, 1OQ8CA-00268, 1OQ8CA-00198, 1OQ8CA-00255, 1OQ8CA-00151, 1OQ8CA-00342, 1OQ8CA-00391, 1OQ8CA-00321, 1OQ8CA-00402, 1OQ8CA-00395, 1OQ8CA-00297, 1OQ8CA-00240, 1OQ89M-00691, 1OQ89M-00690, 1OQ89M-00708, 1OQ89M-00693, 1OQ89M-00687, 1OQ89M-00707, 1OQ89M-00703, 1OQ89M-00679, 1OQ89M-00683, 1OQ89M-00692, 1OQ89M-00610, 1OQ89M-00613, 1OQ89M-00669, 1OQ89M-00635, 1OQ89M-00664, 1OQ89M-00697, P1OQ84F-00080, 1OQ89M-00684, 1OQ89M-00676, 1OQ89M-00695, 1OQ8CA-00086, 1OQ89M-00677, 1OQ8CA-00031, 1OQ8CA-00027, 1OQ8CA-00065, 1OQ8CA-00092, 1OQ8CA-00082, 1OQ89M-00602, 1OQ89M-00680, 1OQ89M-00675, 1OQ89M-00593, 1OQ89M-00686, 1OQ89M-00696, 1OQ89M-00681, 1OQ89M-00704, 1OQ89M-00674, 1OQ89M-00661, 1OQ89M-00663, P1OQ89M-00559, P1OQ89M-00540, 1OQ8CA-00134, 1OQ8CA-00095, 1OQ8CA-00043, 1OQ8CA-00074, 1OQ8CA-00762, 1OQ8CA-00688, 1OQ8CA-00670, 1OQ8CA-00637, 1OQ8CA-00633, 1OQ8E8-00037, 1OQ8E8-00053, 1OQ8CA-00773, 1OQ8CA-00683, 1OQ8CA-00500, 1OQ8CA-00476, 1OQ8CA-00449, 1OQ8CA-00420, 1OQ8CA-00460, 1OQ8CA-00463, 1OQ8CA-00507, 1OQ8CA-00350, 1OQ8CA-00442, 1OQ8CA-00433, 1OQ8CA-00447, 1OQ8CA-00601, 1OQ8CA-00604, 1OQ8CA-00695, 1OQ8CA-00513, 1OQ8CA-00575, 1OQ8CA-00577, 1OQ8CA-00591, 1OQ8CA-00197, 1OQ8CA-00516, 1OQ8CA-00541, 1OQ8CA-00542, 1OQ8CA-00568, 1OQ8CA-00499, 1OQ8CA-00498, P12051M00227, 1OQ8CA-00593, 12946M00532, 12946M00256, 12946M00278, 1OQ7WR-00203
Why was this product recalled?
Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.
What actions did Draeger Medical Systems, Inc. take?
Draeger notified consignees on about 01/14/2025 via Urgent - Medical Device Correction letter. Consignees were instructed to inspect the device to identify if the speaker is configured with the ICS in USB mode or Analog mode. For the speaker set to Analog mode there is an analog cable (aux cable) connected to the backside of the speaker and a light on frontside of the speaker display. If the light is green, there is no issue of speaker audio annunciation. If the light is red, the issue of speaker audio annunciation is present. If an issue with alarm annunciation occurs while operating, follow the steps included in the letter. Ensure all affected users or customers, if further distributed, are notified of the recall notification. They were instructed to complete and return the provided reply card.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1066-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology