muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate Recall
muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate
FDA recall Class II
Ongoing
FDA lists this Class II recall for muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1069-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Apr 22, 2024
- FDA report date
- Feb 5, 2025
- Quantity in commerce
- 62 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate
Affected model & UDI checker
Search the exact identifiers published for Z-1069-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 2050 | muRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate | Check source notice | Not specified |
FDA code information
Soft ware version 1.57.20, UDI: +B595BSM20500/$$7V1.57.20/16D20231129Y
Why was this product recalled?
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
What actions did Murata Vios, Inc. take?
Murata sent an email to its consignees on 04/22/2024 to inform them of the issue and provide the correct workflow to address the issue. The email further requested the posting of the workflow guidance in nurse training areas, add to in-services, and other recurrent training for Vios Monitoring. If the flickering issue is observed on the central station monitor, immediately begin using the bedside tablet for all patient monitoring and alarm management and contact Vios Support at 866-990-1434.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1069-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology