FDA RECALL RECORD · Z-1078-2026

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO) Recall

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

FDA recall Class II

Ongoing

FDA lists this Class II recall for MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO). Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1078-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 10, 2025
FDA report date
Jan 21, 2026
Quantity in commerce
1
Distribution
See key details
FDA cause
Software change control

What is being recalled?

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Affected model & UDI checker

Search the exact identifiers published for Z-1078-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)Check source notice07340201500026
Not specifiedMOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)Check source notice07340201500071
FDA code information

MOSAIQ v 3.10.200, and 3.1.1.0 up to and including v 3.2.3.1 with Particle Base. Software Versions: 3.10.200, 3.1.1.0, 3.1.2.0, 3.1.3.0, 3.1.3.1, 3.1.3.2, 3.1.3.3, 3.1.3.4, 3.2.0.0, 3.2.1.0, 3.2.1.1, 3.2.1.2, 3.2.1.3, 3.2.1.4, 3.2.2.0, 3.2.2.1, 3.2.3.0, 3.2.3.1 UDI-DI: 07340201500026, 07340201500071

View the original FDA recall notice

Why was this product recalled?

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Elekta, Inc. take?

On 12/10/2025 correction notices were emailed to consignees who were asked to do the following: Radiation therapists must always check the meterset values displayed on the MOSAIQ console with the value displayed on the treatment unit console. If a mismatch occurs, follow Caution 9.6 in the MOSAIQ Instructions for Use. Complete and return the acknowledgement form. Firm is working on the correction and will release a Field Safety Modification with the details as soon as possible. Elekta Regional Contact Details: Region America: Application.Support.NA@elekta.com Tel: +1 855 6935358 https://www.elekta.com Region Asia Pacific: HK.TW.Support@elekta.com Tel: + 852 2891 2208 https://www.elekta.com Japan Japan-fco@elekta.com Tel: + 81 3 6722 3800 Fax: +81 3 6436 4231 https://www.elekta.com Region China: FCO.CN@elekta.com Tel : + 86 10 8012 5012 https://www.elekta.com Region Europe: Support.europe@elekta.com Tel : + 46 8 587 254 00 https://www.elekta.com Region Turkey, India & Middle East support.rma@elekta.com Turkey Tel: +90 216 444 6374 India Tel: +1-800-103-7454 Middle East Tel: +00 800 4000 5000 https://www.elekta.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.