FDA RECALL RECORD · Z-1091-2025

Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235 Recall

Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235

FDA recall Class II

Ongoing

FDA lists this Class II recall for Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1091-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 13, 2025
FDA report date
Feb 12, 2025
Quantity in commerce
90 units US; 458 units OUS
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI: (1) 722067 00884838085350(21); (2) 722225 00884838099265(21); (3) 722235 00884838116788(21)

View the original FDA recall notice

Why was this product recalled?

Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?

Philips issued URGENT Medical Device Correction Letter via Certified Mail (USPS) on 1/13/25. Letter states reason for recall, health risk and action to take: " Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. " Follow the additional instructions when positioning the patient table included in Appendix B. These instructions include information of potential finger entrapment areas when moving the table. The information of this Appendix is being included in an Addendum to the Instructions for Use. Additionally, Philips is creating a Quick Reference Card. The Addendum and the Quick Reference Card will be available as from end of March 2025. Please refer to Appendix C on how to download the Addendum to the Instructions for Use and the Quick Reference Card. " In case that the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through the email account: IGT_Recalls@philips.com " Please complete and return the Response form attached to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter and understanding of the issue and required actions to be taken. " Should you experience the issue reported in this letter, please report the event to Philips through your local Philips representative (1- 800-722-9377). Actions planned by Philips IGT-S to correct the problem Philips is releasing an addendum to the instructions for use and a Quick Reference Card that will be available for customers at the end of March 2025. Philips is working on the design of additional brackets for the AD7 tables, and a new redesigned table cover for AD7X tables (ref. FCOs: FCO72200516, FCO72200552, FCO72200553, FCO72200554, FCO72200555, FCO72200556). Philips will contact you to schedule a visit once the ne

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology