The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release. Recall
The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release.
FDA recall Class II
Ongoing
FDA lists this Class II recall for The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1095-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 8, 2025
- FDA report date
- Feb 12, 2025
- Quantity in commerce
- 1351 units
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release.
Affected model & UDI checker
Search the exact identifiers published for Z-1095-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | The Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kit (PN: 71738-001). The Collection Tube (PN: 71550-001) is manufactured by a third-party supplier, MML Diagnostics Packaging, and includes the transport tube, cap, and extraction buffer. Tubes are packaged into PN: 71738-00 at Hologic prior to kit release. | Check source notice | 25420045514765 |
FDA code information
Lot # 901164, 901165, 904850/ UDI: 25420045514765
Why was this product recalled?
Three lots of Rapid fFN Specimen Collection kit (901164, 901165, and 904850) contain Rapid fFN Specimen Collection Tubes which may be empty, underfilled, or overfilled with extraction buffer. The transport tubes are used with the Fetal Fibronectin Enzyme Immunoassay and Rapid fFN¿ for the TLiIQ¿ System which aids in assessing the risk of preterm delivery. The use of empty transport tubes or transport tubes containing insufficient extraction buffer volume may result in invalid results and/or delays in obtaining results. Additionally, the use of underfilled tubes may result in false positive results and the use of overfilled tubes may result in false negative results.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Hologic, Inc. take?
Hologic issued a "Urgent: Medical Device Recall Notification on January 8, 2025 via FedEx. Hologic asked consignees to take the following actions: 1. Please discontinue use of the Rapid fFN Specimen Collection Kit Catalog Number: 71738-001; 2. Please segregate and count how many of the affected Rapid fFN Specimen Collection Kit Catalog Number: 71738-001; 3. Please fill out the Customer Response Form (CRF); 4. Hologic will process and ship the replacement kits as listed on the completed CRF once it has been received; 5. For tests already administered from the affected product lots and if the patient has not given birth, it is at the discretion of the healthcare providers to notify patients who were managed based on a number of clinical decision points including but not exclusive to fetal fibronectin test results. If the patient has given birth, no further action is required.; 6. If you have any questions or concerns regarding this recall notification, please contact Hologic Technical Support
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1095-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology