Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1101-2026

Portex Spinal Tray, Item No. A3729-24 A3595 Recall

Portex Spinal Tray, Item No. A3729-24 A3595

FDA recall Class II

Ongoing

FDA lists this Class II recall for Portex Spinal Tray, Item No. A3729-24 A3595. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1101-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 19, 2025
FDA report date
Jan 21, 2026
Quantity in commerce
64290 total
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Portex Spinal Tray, Item No. A3729-24 A3595

Affected model & UDI checker

Search the exact identifiers published for Z-1101-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedPortex Spinal Tray, Item No. A3729-24 A3595Check source notice30351688075691
Not specifiedPortex Spinal Tray, Item No. A3729-24 A3595Check source notice35019517108211
FDA code information

1. A3729-24 UDI-DI 30351688075691 Lots 6146204 (corrected on 1/22/26, removed "-LSM" suffix. 2. A3595 UDI-DI 35019517108211 Lot 6157980

View the original FDA recall notice

Why was this product recalled?

Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.

What actions did Icu Medical, Inc. take?

On December 19. 2025, the firm issued Urgent Medical Device Correction letters to affected customers. Customers were informed that affected lots may have a crack present on the syringe collar. Customers were instructed to discontinue use of affected lots, and to destroy products following their institution's process for destruction. If destruction is not immediately possible, affected product should be quarantined until disposal.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology