FDA lists this Class II recall for Integra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1103-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 19, 2024
FDA report date
Feb 12, 2025
Quantity in commerce
430 units (43 packs of 10)
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Integra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;

Affected model & UDI checker

Search the exact identifiers published for Z-1103-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
C-CA-23Integra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;Check source noticeNot specified
Not specifiedIntegra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;Check source notice10381780171959
FDA code information

Model: C-CA-23; Item: 33516; UDI-DI: 10381780171959; Lots: C6637.

View the original FDA recall notice

Why was this product recalled?

During the assembly of the cartridge to its connecting body part, the valve may open prematurely resulting in partial or complete emptying of the cartridge before its intended use.

What actions did New Medical Technologies Gmbh take?

On December 19, 2024 "Important Notice Regarding Your Device" letters were sent to customers. Corrective Actions: We have implemented the following corrective and preventive actions to address this issue: 1. Enhanced Testing: All cartridges are now being filled with a minimum specified volume (23.5g tolerance) and subjected to rigorous leakage testing. 2. Updated Specifications: We have revised the interface connection specifications to include precise tolerances and correlated these adjustments with the gas volume requirements. 3. Improved Instructions for Use (IFU): Additional safety precautions regarding gas leakage during assembly are now clearly highlighted in the Instructions for Use (IFU) provided with the product. We are committed to ensuring these actions eliminate the identified problem and prevent its recurrence. Our quality assurance team will also verify the effectiveness of these measures by closely monitoring product performance and feedback. What You Should Do: " Please ensure that all users review the updated Instructions for Use (IFU) to familiarize themselves with the additional safety measures, including warnings regarding the potential for premature gas leakage during cartridge connection. Aligned with that the gas cylinder shall be quickly screwed into the handpiece. " Please ensure that a new O-ring is assembled into the body part of the handpiece We sincerely apologize for any inconvenience this may cause and appreciate your understanding and cooperation. If you have any questions or require assistance, please do not hesitate to contact our customer service team. on January 31, 2025, URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION letters were sent out to sub-accounts from their sole consignee. What You Should Do (Customers (i.e. Medical Facility, etc.)): 1. Please review the notice from UMP (Appendix 1). 2. Please ensure that you review the updated Instructions for Use (IFU) to familiarize yourselves with the additional safety measures, incl

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.