Brand Name: RayStation Product Name: RayStation/RayPlan Recall
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: RayStation Product Name: RayStation/RayPlan. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1110-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Nov 28, 2025
- FDA report date
- Jan 28, 2026
- Quantity in commerce
- 0
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-1110-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 17.0.0 | Brand Name: RayStation Product Name: RayStation/RayPlan | Check source notice | Not specified |
| 17.0.1 | Brand Name: RayStation Product Name: RayStation/RayPlan | Check source notice | Not specified |
| Not specified | Brand Name: RayStation Product Name: RayStation/RayPlan | Check source notice | 07350002010822 |
| Not specified | Brand Name: RayStation Product Name: RayStation/RayPlan | Check source notice | 07350002010945 |
FDA code information
Lot Code: UDI: 0735000201082220250423, 0735000201094520250702 GTIN: 07350002010822, 07350002010945 Serial Numbers: 17.0.0.1270, 17.0.1.113 Software Revision: RayStation v2025, RayStation v2025 SP1 Expiration Date: 2030-10-22
Why was this product recalled?
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did RAYSEARCH LABORATORIES AB take?
RaySearch Laboratories notified consignees via email on about 12/05/2025. Consignees were instructed to follow the outlined steps in the notification letter when adding geometry to an empty ROI with MO or type Bolus, Fixation or Support, educate planning staff and all users about the workaround, inspect units and identify all installed units with the affected software version numbers, and confirm receipt and understanding of the customer notification by responding to the notification email. The issue will be resolved in the next version of RayStation, scheduled for market release in December 2025 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. The upgrade to an unaffected software version will be scheduled in coordination with the authorized personnel, considering the specific circumstances and requirements of each individual case. The upgrade may be conducted remotely or onsite, as deemed appropriate.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1110-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology