Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1112-2025

Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems Recall

Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

FDA recall Class II

Ongoing

FDA lists this Class II recall for Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1112-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 2, 2025
FDA report date
Feb 12, 2025
Quantity in commerce
6,994 US; 10,067 OUS
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

Affected model & UDI checker

Search the exact identifiers published for Z-1112-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
C53987Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay SystemsCheck source noticeNot specified
Not specifiedAccess PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay SystemsCheck source notice15099590736002
FDA code information

UDI-DI 15099590736002 Lots 339062, 439811

View the original FDA recall notice

Why was this product recalled?

Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.

What actions did BECKMAN COULTER, INC. take?

An Urgent Medical Device Recall letter was sent during the week of December 30, 2024, to the affected customers via first class mail and email. If a passing calibration curve can be generated using the Access PCT reagent lots listed above, no action is needed. If repeated calibration failures with a failure code of Max Iterate are obtained using the Access PCT reagent lots listed above, discontinue using the affected reagent lot number.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology