Philips Azurion system; Software Version Number: R3.1; Recall
Philips Azurion system; Software Version Number: R3.1;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Philips Azurion system; Software Version Number: R3.1;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1113-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 12, 2025
- FDA report date
- Jan 28, 2026
- Quantity in commerce
- 228 units (all OUS)
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Philips Azurion system; Software Version Number: R3.1;
Affected model & UDI checker
Search the exact identifiers published for Z-1113-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 722221 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722222 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722223 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722224 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722225 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722226 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722227 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722228 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722229 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722230 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722231 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722232 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722233 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722234 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722235 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| 722236 | Philips Azurion system; Software Version Number: R3.1; | Check source notice | Not specified |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838099203 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838099210 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838099241 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838099258 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838099265 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838099272 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838099227 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838099234 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838116726 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838116733 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838116740 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838116757 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838116764 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838116771 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838116788 |
| Not specified | Philips Azurion system; Software Version Number: R3.1; | Check source notice | 00884838116801 |
FDA code information
Software Version Number: R3.1; (1) Model Number: 722221; System Product Name: Azurion 3 M12; UDI-DI: 00884838099203; Serial Numbers: (2) Model Number: 722222; System Product Name: Azurion 3 M15; UDI-DI: 00884838099210; Serial Numbers: (3) Model Number: 722223; System Product Name: Azurion 7 M12; UDI-DI: 00884838099241; Serial Numbers: (4) Model Number: 722224; System Product Name: Azurion 7 M20; UDI-DI: 00884838099258; Serial Numbers: (5) Model Number: 722225; System Product Name: Azurion 7 B12; UDI-DI: 00884838099265; Serial Numbers: (6) Model Number: 722226; System Product Name: Azurion 7 B20; UDI-DI: 00884838099272; Serial Numbers: (7) Model Number: 722227; System Product Name: Azurion 5 M12; UDI-DI: 00884838099227; Serial Numbers: (8) Model Number: 722228; System Product Name: Azurion 5 M20; UDI-DI: 00884838099234; Serial Numbers: (9) Model Number: 722229; System Product Name: Azurion 3 M12; UDI-DI: 00884838116726; Serial Numbers: 5, 9, 32, 40, 35, 31, 33; (10) Model Number: 722230; System Product Name: Azurion 3 M15; UDI-DI: 00884838116733; Serial Numbers: 25, 54, 50, 52, 55, 30, 63, 60, 67; (11) Model Number: 722231; System Product Name: Azurion 5 M12; UDI-DI: 00884838116740; Serial Numbers: 62, 7, 8, 14, 17, 63, 6, 66, 59, 69, 56, 67, 48, 68, 80, 93, 87, 94, 81, 92, 18; (12) Model Number: 722232; System Product Name: Azurion 5 M20; UDI-DI: 00884838116757; Serial Numbers: 97, 39, 31, 100, 104, 93, 115, 113, 118, 119, 89, 54, 67, 7, 128, 86, 110, 132, 130, 131, 123, 137, 139, 117, 260; (13) Model Number: 722233; System Product Name: Azurion 7 M12; UDI-DI: 00884838116764; Serial Numbers: 239, 100, 233, 240, 176, 227, 223, 130, 161, 194, 243, 56, 193, 280, 75, 253, 156, 288, 72, 281, 304, 300, 105, 295, 312, 310, 285, 292, 290, 298, 297; (14) Model Number: 722234; System Product Name: Azurion 7 M20; UDI-DI: 00884838116771; Serial Numbers: 731, 727, 261, 473, 683, 722, 562, 750, 745, 766, 265, 701, 400, 310, 421, 314, 182, 579, 96, 763, 272, 612, 15, 844, 105, 702, 751, 384, 851, 869, 743, 337, 651, 455, 706, 312, 563, 709, 764, 871, 846, 278, 713, 802, 868, 277, 742, 184, 819, 759, 910, 928, 415, 414, 945, 961, 355, 933, 692, 920, 569, 939, 909, 553, 837, 856, 571, 756, 941, 570, 761, 981, 353, 888, 964, 820, 746, 736, 876, 599, 1017, 908, 487, 983, 495; (15) Model Number: 722235; System Product Name: Azurion 7 B12; UDI-DI: 00884838116788; Serial Numbers: 96, 79, 111, 100, 101, 38, 27, 92, 78, 118, 48, 131, 135, 22, 125, 103, 138, 139, 137; (16) Model Number: 722236; System Product Name: Azurion 7 B20; UDI-DI: 00884838116801; Serial Numbers: 241, 297, 198, 215, 305, 274, 299, 78, 279, 287, 318, 353, 324, 331, 293, 245, 314, 310, 317, 312, 77, 358, 206, 383, 380, 370, 377, 385, 405, 379, 328;
Why was this product recalled?
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?
On December 12, 2025, Philips sent URGENT Medical Device Correction letters to customers. Actions to be taken: - Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issues. " In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. " Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. " Complete and return the response form included in this Urgent Medical Device Correction letter to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issues and required actions to be taken. " If you experience an issue described in this letter, please report it to your local Philips representative. Your local Philips representative will contact you to schedule visits to install the software updates once available. ***Update May 8, 2026*** Previously, Philips planned to resolve issue 1 with software update 3.1.5 and issue 2 with software update 3.1.15. Philips has now updated their plan thusly: Both issues 1 and 2 will be included with software release 3.1.15. Anticipated release date of software version 3.1.15 has been moved to Q3 2026.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1113-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology