Mazor X robotic guidance system REF: TPL0059 Recall
Mazor X robotic guidance system REF: TPL0059
FDA recall Class II
Ongoing
FDA lists this Class II recall for Mazor X robotic guidance system REF: TPL0059. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1122-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 10, 2025
- FDA report date
- Jan 28, 2026
- Quantity in commerce
- 549 systems
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Mazor X robotic guidance system REF: TPL0059
Affected model & UDI checker
Search the exact identifiers published for Z-1122-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| TPL0059 | Mazor X robotic guidance system REF: TPL0059 | Check source notice | Not specified |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 07290109180465 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 07290109184524 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 07290109181158 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 00763000635169 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 07290109183213 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 07290109184517 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 00763000419134 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 00763000431761 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 07290109184098 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 07290109184838 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 07290115751376 |
| Not specified | Mazor X robotic guidance system REF: TPL0059 | Check source notice | 07290115751895 |
FDA code information
Software versions: 5.0.1, 5.1.2, 5.1.3/ UDI: 07290109180465, 07290109184524, 07290109181158, 00763000635169, 07290109183213, 07290109184517, 00763000419134, 00763000431761, 07290109184098, 07290109184838, 07290115751376, 07290115751895
Why was this product recalled?
Software errors that can result in incorrect surgical instrument positioning during spinal surgery.
What actions did Mazor Robotics Ltd take?
Beginning December 10, 2025, Medtronic (Mazor Robotics, Ltd) issued a Urgent Medical Device Recall Notification to affected consignees via E-Mail. Medtronic asked consignees to take the following actions: 1. Please review this information with all physician users and/or post a copy of this notification with your Mazor X system until the software update is completed. 2. Please confirm via the enclosed confirmation form that this notification has been communicated within your facility with all physician users. Send the completed Customer Confirmation Form to Medtronic via email at neuro.quality@medtronic.com. 3. Provide this notification to those who need to be aware within your organization or to any organization where the affected systems have been transferred. 4. Maintain a copy of this notice in your records. 5. Continue to use the Mazor X system as instructed in the user manual and with the mitigations provided below in Appendix A.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1122-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology