FDA RECALL RECORD · Z-1127-2026

Merit Medical , Custom Manifold Kit REF: K09-13203A Recall

Merit Medical , Custom Manifold Kit REF: K09-13203A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Merit Medical , Custom Manifold Kit REF: K09-13203A. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1127-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 4, 2025
FDA report date
Jan 28, 2026
Quantity in commerce
2192 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Merit Medical , Custom Manifold Kit REF: K09-13203A

Affected model & UDI checker

Search the exact identifiers published for Z-1127-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
K09-13203AMerit Medical , Custom Manifold Kit REF: K09-13203ACheck source noticeNot specified
Not specifiedMerit Medical , Custom Manifold Kit REF: K09-13203ACheck source notice00884450479964
FDA code information

K09-13203A: UDI: 00884450479964 /Lot: H2671718 H2681531 H2758436 H2792533 H2796074 H3046979 H3052442 H3147779 H3246149 H3254530 H3297407

View the original FDA recall notice

Why was this product recalled?

Inflation device handle may detach from the syringe during procedure.

What actions did Merit Medical Systems, Inc. take?

On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at RESPONSE@merit.com within seven (7) calendar days

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history