FDA RECALL RECORD · Z-1128-2026

Merit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922B Recall

Merit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922B

FDA recall Class II

Ongoing

FDA lists this Class II recall for Merit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922B. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1128-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 4, 2025
FDA report date
Jan 28, 2026
Quantity in commerce
Not reported
Distribution
See key details
FDA cause
Process control

What is being recalled?

Merit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922B

Affected model & UDI checker

Search the exact identifiers published for Z-1128-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMerit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922BCheck source notice00884450389683
Not specifiedMerit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922BCheck source notice00884450264003
Not specifiedMerit Medical , Custom Procedure Kit Reference Numbers: K12T-04270A K12T-07956 K12T-10922BCheck source notice00884450481660
FDA code information

Reference Number/UDI Number/Lot Number(s) K12T-04270A: UDI: 00884450389683/ Lot: T2643938 T2728112 T2783917 T2806210 T2811252 K12T-07956: UDI: 00884450264003/ Lot: K2618155 K2618156 K12T-10922B: UDI: 00884450481660/Lot: T2851185 T2876341 T2990792 T3063197 T3070788 T3165727 T3223759 T3312055

View the original FDA recall notice

Why was this product recalled?

Inflation device handle may detach from the syringe during procedure.

What actions did Merit Medical Systems, Inc. take?

On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at RESPONSE@merit.com within seven (7) calendar days

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.