Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/14Shaft/Left/Std/186mm Recall
Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/14Shaft/Left/Std/186mm Model/Catalog Number: 02.426.691S Product Description: 2 Column Distal Radius Plate - 6 Head, 14 Shaft, Standard Left - 186mm Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/14Shaft/Left/Std/186mm. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1132-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 19, 2025
- FDA report date
- Jan 28, 2026
- Quantity in commerce
- 64 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/14Shaft/Left/Std/186mm Model/Catalog Number: 02.426.691S Product Description: 2 Column Distal Radius Plate - 6 Head, 14 Shaft, Standard Left - 186mm Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-1132-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 02.426.691S | Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/14Shaft/Left/Std/186mm | Check source notice | Not specified |
| Not specified | Brand Name: VOLT" Wrist Treatment System Product Name: 2.4/2.7 2Column Dist Radius Plate 6Hd/14Shaft/Left/Std/186mm | Check source notice | 00196449020522 |
FDA code information
Model No: 02.426.691S; UDI-DI: 00196449020522; Lot Number: 291293
Why was this product recalled?
The supplier manufactured anatomical left plates with an incorrect thread orientation.
What actions did Tyber Medical take?
On December 19, 2025, an URGENT MEDICAL DEVICE REMOVAL NOTIFICATION email was sent to their only customer. Actions to be taken: 1. Acknowledge the receipt of this communication via email correspondence to WeCare@exalta.com 2. Further distribute this communication through notifying the accounts that have received the identified devices. 3. Identify via product label and quarantine the impacted devices under DePuy Synthes control all devices that have not been distributed for sale. An RMA will be provided for the return of goods to Tyber Medical. 4. Identify via product label and quarantine the impacted devices that have been distributed. An RMA will be provided for the return of goods to Tyber Medical. 5. Complete the customer response form following this notification, or provide written notification, and return it to Tyber Medical via email correspondence to WeCare@exalta.com. A sub-notification letter was also provided to the distributor to provide to customers with the following actions: 1. Identify via product label and quarantine the impacted devices in your possession. 2. Complete the subsequent Acknowledgement and Receipt From and return it to DePuy Synthes via email correspondence, with Tyber Medical copied at WeCare@exalta.com 3. Return the impacted product to DePuy Synthes for replacement.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1132-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology