FDA RECALL RECORD · Z-1156-2026

Patient Information Center iX; Software Version Number: 4.5.0 Recall

Patient Information Center iX; Software Version Number: 4.5.0

FDA recall Class II

Ongoing

FDA lists this Class II recall for Patient Information Center iX; Software Version Number: 4.5.0. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1156-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 22, 2025
FDA report date
Jan 28, 2026
Quantity in commerce
146 units (41 US, 105 OUS)
Distribution
See key details
FDA cause
Software Design Change

What is being recalled?

Patient Information Center iX; Software Version Number: 4.5.0

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Model Numbers: (1) 866389, (2) 866390; UDI-DI: (1) 00884838127074, (2) N/A; Serial Numbers: 4Z24-5CKL-P, 0P6C-4TL3-D, 6J58-3LL6-0, 216G-7PGV-E, 0F2U-1BLC-H, 291E-7HJU-A, 1M6W-1UP5-1, 4620-0XNW-5, 4W5R-3FHA-L, 6B73-3ZKP-K, 3E79-49M4-3, 4F6Y-70HU-P, 0128-2WG5-0, 0830-3RLW-W, 097K-7LPC-5, 0D2V-1BLC-8, 120T-17M9-A, 173C-20JM-J, 190B-64MU-U, 1K2J-15LB-P, 1L3T-7KGB-E, 2K06-4ZPR-X, 3916-5DJ6-Z, 3V7H-06MT-E, 412A-2WG5-Y, 4271-2NL7-M, 4Z3L-5PP3-0, 5D0E-3EH3-3, 5E0B-3EH3-D, 5E25-14LG-9, 657X-31N5-L, 6G5U-3FHA-R, 781T-7TJA-B, 7N3U-51MJ-B, 7P31-7NPM-P, 283X-7ZPB-A, 7H77-09GU-Z, 4C59-26K6-P, 4N0L-0FKE-2, 5Y0A-02P6-L, 494H-1CJJ-3, 326G-7PGV-A, 292G-3YNZ-9, 0V2A-6DHP-0, 6354-58M8-C, 4W5B-6JK9-A, 5M7K-0VKT-T, 3401-1FHE-M, 6R30-5BNT-N, 6N2E-7YK1-1, 6R0H-6AMX-T, 107C-62PF-Y, 2X1T-7TJA-H, 1M6G-7PGV-K, 1C4Z-6KKD-H, 0B3Z-2RHR-L, 7H39-3JGK-U, 0M0V-6FKB-5, 793D-41PZ-X, 233G-46MR-V, 782Z-3RLW-D, 5N2B-5YJP-G, 0L5F-0ZNF-J, 4G4N-2LKY-K, 2M48-4XM7-Y, 133W-3UN7-B, 295N-0YG8-Y, 5D7W-00G0-1, 5R78-6CPV-0, 3C79-49M4-9, 3T24-5CKL-A, 6H74-27LY-T, 186W-1UP5-8, 0021-1DNN-P, 0247-7ELE-R, 0D5Y-32PN-A, 0J7W-31N5-2, 0R2W-1VNP-P, 0Y5Z-7FHX-N, 121L-1KG1-M, 1366-7XLV-X, 144W-7RPJ-5, 1472-5WKF-X, 1C19-1MHK-F, 1G0F-69LP-G, 1J5U-5HMV-N, 1P4F-34MB-W, 1P5C-6JK9-D, 212R-6HHJ-P, 221V-7TJA-4, 227N-7JG6-J, 232W-7DL9-G, 2A5X-2PJT-0, 2A75-27LY-3, 2F5J-0JLM-7, 2J19-1MHK-W, 2L16-59N8-M, 2M73-5WKF-U, 2N3V-6YM2-X, 2P6J-7PGV-E, 2Y5M-5AGH-E, 326W-21KC-B, 332M-47P7-Y, 3A29-0CKW-3, 3C0U-7AHH-6, 3C2U-05HF-W, 3E05-4ZPR-U, 3G6H-39G4-Y, 3H1T-22HW-5, 3H5R-5HMV-M, 3H79-09GU-5, 3J2K-2TP9-1, 3W1Y-3CH6-J, 4L7Y-00G0-F, 4M77-1HM0-G, 4N3D-20JM-K, 4R0G-60JV-1, 4V2Z-7NPM-M, 4W08-0UH7-U, 4X6C-1PJ5-J, 4Y08-0UH7-L, 500W-6WNX-A, 594D-0EG2-9, 5U0A-3WPK-C, 5W1G-50P2-N, 5Z6J-3HL1-B, 617W-31N5-F, 637V-7BG9-W, 673T-67ME-B, 686N-3NNV-6, 6D3P-4WMG-A, 6K70-3XNY-W, 6P42-35J4-V, 6U1T-75JN-Y, 771A-2JPY-0, 7F5L-73HZ-3, 7G3A-13JG-L, 7K0P-4AH2-6, 7Z2R-05HF-X, 1J1P-6PKN-T, 4R7H-7LPC-J, 282N-0WPA-X, 5G35-04K8-E, 366L-2XHV-A, 5L7D-2UGW-T, 4N0L-0FKE-2;

View the original FDA recall notice

Why was this product recalled?

A recent software (SW) patch modifies Mobile Event Notification filter settings without providing any indication to the user when upgrading the system.

What actions did Philips North America take?

On or about December 22, 2025 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: Review the Mobile Notification Event Filters settings and confirm all the events intended to be sent to mobile devices automatically are enabled for Mobile Notification Event Filters. " Pass this notice to all those who need to be aware within your organization or to any organization where affected product(s) have been potentially transferred. Actions planned by Philips to correct the problem " Philips representative will contact you to obtain PIC iX 4.5.1 SW patch for PIC iX systems running software version 4.5.0. If you need any further information, please contact your local Philips representative, or call (800) 722-9377.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.