*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530; Recall
*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;
FDA recall Class I
Ongoing
FDA lists this Class I recall for *HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1162-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Dec 19, 2025
- FDA report date
- Feb 11, 2026
- Quantity in commerce
- 150 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;
Affected model & UDI checker
Search the exact identifiers published for Z-1162-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | *HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530; | Check source notice | 8714729904557 |
FDA code information
Material Number (UPN): M00553530; UDI-DI: 8714729904557; Lot Numbers: 35985705, 35986358, 36147098, 36147222, 36165642, 36176423, 36190575, 36191222, 36194448, 36218745, 36227778, 36227779, 36241500, 36271259, 36280398, 36293880, 36300839, 36301620, 36343103, 36343104, 36364646, 36425887, 36469811, 36471228, 36491879, 36492660, 36506029, 36510220, 36510221, 36513401, 36513402, 36513403, 36517139, 36517220, 36525208, 36525209, 36525210, 36534514, 36534518, 36536259, 36536480, 36536481, 36551000, 36551001, 36642564, 36642565, 36642756, 36645775, 36663654, 36677842, 36677843, 36682759, 36684887, 36684888, 36730665, 36730666, 36747824, 36747825, 36753898, 36753899, 36757493, 36757494, 36761827, 36766370, 36766371, 36783265, 36792467, 36792468, 36792469, 36792470, 36792471, 36799315, 36815090, 36815092, 36832774, 36833683, 36836357, 36844783, 36845422, 36846169, 36846176, 36851617, 36851792, 36853596, 36865505, 36873631, 36873632, 36873906, 36874863, 36904228, 36904229, 36904230, 36907002, 36907003, 36907004, 36907028, 37102533, 37102534, 37102535, 37102536, 37102537, 37459567, 37459568, 37459569, 37459570, 37463133, 37470670, 37498192, 37511243, 37519095, 37519096, 37844029, 37519097, 37520927, 37530320, 37530321, 37530333, 37533308, 37541933, 37571927, 37571928, 37579364, 37579366, 37579367, 37580630, 37585525, 37585526, 37589261, 37589262, 37594668, 37649481, 37649482, 37649483, 37650631, 37694052, 37699011, 37699012, 37708202, 37779125, 37779126, 37780822, 37780823, 37780824, 37823589, 37823590, 37827568, 37827569, 37833165, 37833176, 37834499;
Why was this product recalled?
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
What actions did BOSTON SCIENTIFIC CORPORATION take?
On December 19, 2025, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: - Immediately stop further distribution or use of any remaining affected devices. Remove them from your facility s inventory and segregate them in a secure location until they can be returned to Boston Scientific. " Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. " Share this notice with any health care professional within your organization and with any organization where the affected products may have been transferred. " If you are a distributor, this notification must be forwarded to your customers to ensure notification of this product removal is carried out to the end-user level. " If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. " Boston Scientific reminds healthcare professionals that AXIOS Stents should only be used as indicated in Instructions For Use.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1162-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.