Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1162-2026

*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530; Recall

*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;

FDA recall Class I

Ongoing

FDA lists this Class I recall for *HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1162-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Dec 19, 2025
FDA report date
Feb 11, 2026
Quantity in commerce
150 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;

Affected model & UDI checker

Search the exact identifiers published for Z-1162-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specified*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;Check source notice8714729904557
FDA code information

Material Number (UPN): M00553530; UDI-DI: 8714729904557; Lot Numbers: 35985705, 35986358, 36147098, 36147222, 36165642, 36176423, 36190575, 36191222, 36194448, 36218745, 36227778, 36227779, 36241500, 36271259, 36280398, 36293880, 36300839, 36301620, 36343103, 36343104, 36364646, 36425887, 36469811, 36471228, 36491879, 36492660, 36506029, 36510220, 36510221, 36513401, 36513402, 36513403, 36517139, 36517220, 36525208, 36525209, 36525210, 36534514, 36534518, 36536259, 36536480, 36536481, 36551000, 36551001, 36642564, 36642565, 36642756, 36645775, 36663654, 36677842, 36677843, 36682759, 36684887, 36684888, 36730665, 36730666, 36747824, 36747825, 36753898, 36753899, 36757493, 36757494, 36761827, 36766370, 36766371, 36783265, 36792467, 36792468, 36792469, 36792470, 36792471, 36799315, 36815090, 36815092, 36832774, 36833683, 36836357, 36844783, 36845422, 36846169, 36846176, 36851617, 36851792, 36853596, 36865505, 36873631, 36873632, 36873906, 36874863, 36904228, 36904229, 36904230, 36907002, 36907003, 36907004, 36907028, 37102533, 37102534, 37102535, 37102536, 37102537, 37459567, 37459568, 37459569, 37459570, 37463133, 37470670, 37498192, 37511243, 37519095, 37519096, 37844029, 37519097, 37520927, 37530320, 37530321, 37530333, 37533308, 37541933, 37571927, 37571928, 37579364, 37579366, 37579367, 37580630, 37585525, 37585526, 37589261, 37589262, 37594668, 37649481, 37649482, 37649483, 37650631, 37694052, 37699011, 37699012, 37708202, 37779125, 37779126, 37780822, 37780823, 37780824, 37823589, 37823590, 37827568, 37827569, 37833165, 37833176, 37834499;

View the original FDA recall notice

Why was this product recalled?

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

What actions did BOSTON SCIENTIFIC CORPORATION take?

On December 19, 2025, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: - Immediately stop further distribution or use of any remaining affected devices. Remove them from your facility s inventory and segregate them in a secure location until they can be returned to Boston Scientific. " Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device. " Share this notice with any health care professional within your organization and with any organization where the affected products may have been transferred. " If you are a distributor, this notification must be forwarded to your customers to ensure notification of this product removal is carried out to the end-user level. " If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. " Boston Scientific reminds healthcare professionals that AXIOS Stents should only be used as indicated in Instructions For Use.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.