Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20 and 1.56.00 Recall
Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20 and 1.56.00
FDA recall Class II
Ongoing
FDA lists this Class II recall for Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20 and 1.56.00. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1169-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jun 6, 2023
- FDA report date
- Feb 19, 2025
- Quantity in commerce
- 44 systems
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20 and 1.56.00
Affected model & UDI checker
Search the exact identifiers published for Z-1169-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| BSM2050 | Vios Monitoring System Bedside Monitor, Model BSM2050, Software Version Numbers 1.54.20 and 1.56.00 | Check source notice | Not specified |
FDA code information
All systems with Software Version Numbers 1.54.20 and 1.56.00 affected; UDI-DI: B595BSM20500
Why was this product recalled?
A premature failure condition was identified with the L2050R pulse oximeter SpO2 sensor that connects to the L2050G Vios chest sensor device. When a failure occurs, a Pulse Ox Sensor Off warning condition will be incorrectly displayed by the system.
What actions did Murata Vios, Inc. take?
A notification letter dated 6/6/23 was sent to customers. The purpose of this letter is to advise you of a premature failure condition that has been identified with the L2050R pulse oximeter SpO2 sensor that connects to the L2050G Vios chest sensor device. The root cause of the underlying issue has been determined to be due to the flex circuitry degrading more rapidly than previously characterized. When a failure occurs, a Pulse Ox Sensor Off warning condition will be incorrectly displayed by the system. All SpO2 pulse oximeter sensor model # L2050R have the potential for this identified early failure after approximately 3-4 months of persistent usage. Failure analysis has determined that the flexible circuitry that connects the emitter diode to the light sensor can degrade, causing an open circuit condition to exist. When the SpO2 failure condition is identified, the open circuit condition can be incorrectly reported as a sensor off condition via a technical alarm, and the SpO2 measurement will no longer be provided. This error condition has not resulted in any injury or harm to monitored patients. SpO2 sensor off condition does not affect the ability for the chest sensor to correctly monitor other vital signs. If the noted condition is observed, the SpO2 sensor can be safely unplugged from the distal end of the sensor, near the chest sensor. The sensor should be removed from inventory, properly identified, and returned for analysis. To initiate an RMA, please call customer service at +1-866-990-1434. Murata Vios has identified the root cause of the failure and is in the process of taking the following actions: - Evaluation of alternative reinforced flexible circuit designs that can withstand longer use cycles. - Evaluation of adding semi-rigid material around flexible circuit to minimize stresses due to bending. - Identification of software changes that can correctly identify and discern circuit failure vs. sensor off and provide the appropriate warning mes
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1169-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology