Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1178-2025

AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700 Recall

AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700,

FDA recall Class II

Ongoing

FDA lists this Class II recall for AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1178-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 23, 2024
FDA report date
Feb 26, 2025
Quantity in commerce
701 box / 8,412 eaches
Distribution
See key details
FDA cause
Process change control

What is being recalled?

AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700,

Affected model & UDI checker

Search the exact identifiers published for Z-1178-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
9700AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700Check source noticeNot specified
Not specifiedAnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700Check source notice00610075095810
FDA code information

UDI/DI 00610075095810, Lot Numbers: 4J162, 4J172, 4J182, 4L172, 4L182

View the original FDA recall notice

Why was this product recalled?

Hollister received reports of decreased skin barrier wear time which could can lead to tube migration

What actions did Hollister Incorporated take?

Hollister issued an URGENT - MEDICAL DEVICE RECALL notice to its consignees on 12/23/2024 via FedEx and email. The notice explained the reason for the recall and potential risk and instructed customers to quarantine affected inventory, discard the product per their local procedures for product destruction, and Please complete and return the response form attached as soon as possible, even if you do not have affected product in your possession and send to ComplaintCommunication@Hollister.com. For product credit requests: If you are a direct customer of Hollister, contact Hollister Incorporated Customer Service for product credit at: 1-800-323-4060 (prompt #1). If you purchase product from a distributor, please contact your distributor for product credit. Hollister has posted this recall on their website with the link to the page.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.