Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1179-2026

Alphenix INFX-8000V, interventional fluoroscopic x-ray system Recall

Alphenix INFX-8000V, interventional fluoroscopic x-ray system

FDA recall Class II

Ongoing

FDA lists this Class II recall for Alphenix INFX-8000V, interventional fluoroscopic x-ray system. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1179-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 19, 2025
FDA report date
Feb 4, 2026
Quantity in commerce
332 units
Distribution
See key details
FDA cause
Process design

What is being recalled?

Alphenix INFX-8000V, interventional fluoroscopic x-ray system

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

serial numbers: B2A1522117, E0514103, B2A15X2238, B2A1792589, A2A2242418, A2A21X2345, B2A1692413, A2A2152263, A2A2262450, A2A22Z2576, W4A1252250, A2A20Z2167, W4A10X2001, A2A2412831, W1C0552181, W1D0842501, W4D13Z2452, W4A1322350, W4D1422467, B2A1472002, A1C1982123, W2A1062134, W1B08X2042, W2A1072139, W4A1242220, A2A2382741, B2A17Z2627, A2A2152264, W4C1362387, W4D1422465, D4W1262257, W1D0892543, W4D1462478, W4A1182118, W4A11X2133, W4A11Z2167, N/A Deinstalled, A1B18Z7003, W4A1252243, DFP800A-0001, A3552017, W2A1092165, W1D0882537, A2A2122203, W4A10Z2022, W4B1332361, A2A23Y2788, W4A12Z2328, B2A1662374, W4A12Z2329, B2A1622309, B2A14X2056, A2A2272459, A2A2172291, W4A1132061, W1D0832488, W1D0882536, W2A0962022, W2A0982038, W2A0972026, B2A1642351, A2A2152255, W1D0752392, A2A2182308, A2A2412836, A2A2112189, A2A21X2336, A1C19Y2164, A1C1942056, A2A2192328, A2A2332643, W4C1372398, A2A22Y2553, A2A2132226, W4A1262256, B2A1472013, W2A1022097, W2A1022094, W4A10Z2017, W1B08Y2055, B2A1492051, B2A1492052, W4A10Z2021, W4A12Z2332, W4D1412463, W4A1122040, B2B1862727, B2B1812651, B2A1662376, B2A16X2448, B2A1742547, W4A1242230, W4A1272271, G3514035, B2A17X2601, W1D07Z2454, W4A1272272, A2A2142238, B2B1862746, B2B18X2800, A2A2022014, A2A2032029, A2A2172288, A2A2122199, A2A2152258, A2A2012004, W4A10X2004, B2B18X2810, B2B1952847, A2A2142244, A2A2172297, W2A0992045, W4A1262253, W4A1282278, W4A12Z2330, W4A1312340, W2A09Z2075, B2A1652363, W4A1272266, W4A1162079, A2A2152253, W2A1092160, W4D1452476, W1D07Y2445, W2A1072147, A1C1962090, A1C1962089, B2A1652362, W4D14Y2481, A2A20Z2157, A2A22Y2547, A2A2162266, W4A1212183, W4A1212189, W4A1212190, W4A11Z2172, W4A1212178, W4A1212191, W4A1212185, W4A1212184, W4A1132060, W4A1242222, B2A1612295, B2A16Y2468, W4A1222003, A4512077, W2A0982043, W1D0892547, W2A0972030, W2A1042109, A2A2322602, W2A1047001, A1C1982121, A2A2332647, B2A16Y2455, B2A16Y2456, B2A17X2608, B2A17X2605, B2A17X2607, B2A1592231, A2A2232398, B2B1822666, W1B08Z2064, W1B08Z2067, A1C1942063, A1B1922042, B2B1842698, B2B1882777, B2B1892792, W4A1162083, A1C1962094, W4A1222202, W4A12X2299, B2A15X2243, A2A20Y2153, W2A1072141, W4A10Z2014, W4A1142064, W4A1252239, W2A1062136, A2A21X2344, W1D0697017, A2A2022026, A2A22Z2572, W1D0882538, A2A2142237, A2A20Z2161, A1C1992148, W4A1252241, B2A16X2454, W4A1242223, W4A1162089, A2A2242410, B2A1622320, W1D0812470, A2A2182307, A2A2332638, W4A1272240, B2A16Y2474, A2A2262444, A1C1952074, B2A1612300, W4A10Y2006, B2A1582203, W4A1242231, W4A12X2300, B2A1522121, B2A1532152, A2A2142245, B2A1512107, W4D1452473, A1C1992135, A2A2132229, A2A2092109, A2A2052055, A2A2142241, W2A09Y2065, A2A2062064, A2A21X2347, W4C1382414, B2A16Z2492, B2B1582210, B2A14X2062, B2A14Z2097, B2B1882769, W2A0962024, W4A1272261, W2A1052126, W2A09Y2064, B2A1622321, W4A10Z2018, W4A1262254, W4A1182121, W4C13X2433, W4C13X2432, W4C1372401, W1C0632285, A2A2282476, W4C1362389, W4A1322353, B2A15Y2252, W2A1042114, W2A1052125, W2A1072142, W4A1312343, W4A1142063, W4C1362388, W4C1352376, W4C1352378, B2B1862733, A1C1962097, B2B2032825, B2A1592224, A1C1962088, W1B08X2026, W1D0872529, W1D0732379, B2B1872763, B2A17Z2630, B2A17Z2629, W4A11X2132, W4A1112033, F1514114, W4A1122043, W4A1242237, W1D0697015, W1B08X2036, A2A2132227, W1B08Y2050, W4A1112035, W2A09Z2078, A2A2292515, A2A2422852, B2A1582215, W4A1272262, A1C1942064, W4C1342371, A1C19X2155, W4A1212186, W4A1212175, A2A2072088, B2A16Y2467, B2A1612293, B2B18Z2830, B2B1812654, W4A11Z2169, W4A1222196, W4A1212174, W1D0862519, W4A1217002, W4A1242225, B2A14X2057, A2A2432873, W1D0892545, A2A2232396, A2A2382740, W4D1452474, B2A17Z2634, B2B1842701, W2A1072145, F9624025, B2A1532147, A2A2262447, A2A2142246, B2A14Z2099, A2A2152262, B2B1832680, A2A2242429, W1D0782416, B2B1862728, B2B1862735, B2A1642352, B2A16Z2480, W4A1312344, A2A2082098, W2A0982041, B2A1772578, A2A2092106, W4A1162085, B2B1832692, A2A2142236, B2A1552177, A1B1932051, A1B1932049, B2A1682401, W4A12Z2323, B2A1712502, B2A17X2604, W4A12Y2316, B2A1712504, B2A1512106, A2A20X2083.

View the original FDA recall notice

Why was this product recalled?

It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed

What actions did Canon Medical System, USA, INC. take?

Canon Medical Systems USA issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 12/19/2025 via letter. The notice explained the issue, risk to health, and requested the following: "Request to customers: Should the suspended device make an abnormal noise, or a sensor error is displayed before the corrective action is implemented on your system, please stop using the device and call your local Canon Service Team. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at intouchdispatch@us.medical.canon."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology