Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1197-2025

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia Recall

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia

FDA recall Class II

Ongoing

FDA lists this Class II recall for Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1197-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 16, 2024
FDA report date
Feb 26, 2025
Quantity in commerce
4 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia

Affected model & UDI checker

Search the exact identifiers published for Z-1197-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
US501Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermiaCheck source noticeNot specified
Not specifiedHot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermiaCheck source notice00855913001817
FDA code information

UDI-DI UDI-DI 00855913001817, Serial Numbers: 1304292, 1304293, 1304296, 1304297

View the original FDA recall notice

Why was this product recalled?

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

What actions did Augustine Temperature Management, LLC take?

Augustine Temperature Management began notifying its consignees on 12/16/2024 via email and visit to exchange the affected product.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.