Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1210-2025

A.L.P.S. mvX -ANATOMIC LAT FIB PLATE 4H RT: Anatomic Lateral Fibula Plate, 4-hole, Right Recall

A.L.P.S. mvX -ANATOMIC LAT FIB PLATE 4H RT: Anatomic Lateral Fibula Plate, 4-hole, Right Model/Catalog Number: 770708042

FDA recall Class II

Ongoing

FDA lists this Class II recall for A.L.P.S. mvX -ANATOMIC LAT FIB PLATE 4H RT: Anatomic Lateral Fibula Plate, 4-hole, Right. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1210-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 21, 2025
FDA report date
Mar 5, 2025
Quantity in commerce
397
Distribution
See key details
FDA cause
Process control

What is being recalled?

A.L.P.S. mvX -ANATOMIC LAT FIB PLATE 4H RT: Anatomic Lateral Fibula Plate, 4-hole, Right Model/Catalog Number: 770708042

Affected model & UDI checker

Search the exact identifiers published for Z-1210-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
770708042A.L.P.S. mvX -ANATOMIC LAT FIB PLATE 4H RT: Anatomic Lateral Fibula Plate, 4-hole, RightCheck source noticeNot specified
Not specifiedA.L.P.S. mvX -ANATOMIC LAT FIB PLATE 4H RT: Anatomic Lateral Fibula Plate, 4-hole, RightCheck source notice00196449015535
FDA code information

UDI-DI 00196449015535; Lot Numbers 267845, 261287, 265780, 24033BU01

View the original FDA recall notice

Why was this product recalled?

Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices

What actions did Tyber Medical take?

Tyber Medical issued Urgent Medical Device Recall letter on 1/24/25 via email to the Distributor. Letter states reason for recall, health risk and action to take: 1. Acknowledge the receipt of this communication via email correspondence to feedback@tybermed.com. 2. Further distribute this communication to all applicable parties and/or customers that possess the impacted inventory. 3. Identify via laser-marked lot number and quarantine the impacted plates under Zimmer Biomet s control that have not been distributed for sale. An RMA will be provided for the return of goods to Tyber Medical for further inspection. 4. Identify via laser-marked lot number and quarantine the impacted plates that have been distributed. An RMA will be provided for the return of goods to Tyber Medical for further inspection. 5. Complete the customer response form located in Appendix 2 and return it to Tyber Medical via email correspondence to feedback@tybermed.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology