MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1. Recall
MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1.
FDA recall Class II
Ongoing
FDA lists this Class II recall for MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1213-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 3, 2025
- FDA report date
- Feb 4, 2026
- Quantity in commerce
- 9 units
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1.
Affected model & UDI checker
Search the exact identifiers published for Z-1213-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1. | Check source notice | 00884838104112 |
| Not specified | MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1. | Check source notice | 00884838112858 |
FDA code information
MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 782120; UDI-DI: 00884838104112; Serial Numbers: 65019, 65060, 65066, 65065, 65051, 65038, 65005. (2) Product Code (REF): 782153; UDI-DI: 00884838112858; Serial Numbers: 68010, 68017.
Why was this product recalled?
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Philips North America take?
Philips began notification of consignees on about 12/03/2025 via mailed Urgent Medical Device Correction letter, with a corrected notification sent on about 12/18/2025. Consignees were instructed to identify if any affected units were on hand, that the systems could continue to be use in accordance with their intended use. They were asked to provide the notification to all users of the device for awareness, maintain the Urgent Medical Device Correction letter with affected systems until the software upgrade is installed, and complete and return the provided response form. A Philips representative will contact each consignee to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the issues beginning in January 2026 (reference FCO78100585, FCO78100620).
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1213-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology