LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest. Recall
LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.
FDA recall Class II
Ongoing
FDA lists this Class II recall for LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1232-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 3, 2025
- FDA report date
- Mar 5, 2025
- Quantity in commerce
- 1
- Distribution
- See key details
- FDA cause
- Employee error
What is being recalled?
LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.
Affected model & UDI checker
Search the exact identifiers published for Z-1232-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 99507-000001 | LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest. | Check source notice | Not specified |
| Not specified | LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest. | Check source notice | 00883873813751 |
FDA code information
Model/Catalog Number: 99507-000001 UDI-DI code: 00883873813751 Serial Number: 38289467
Why was this product recalled?
Incorrect keypad was incorrectly installed onto a defibrillator/monitor.
What actions did Physio-Control, Inc. take?
On 02/03/2025, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via mail to a customer inform them that Stryker discovered that a keypad was incorrectly installed onto a LIFEPAK 20e defibrillator/monitor. There is a possibility that the incorrectly installed keypad may cause difficulty or the inability to actuate the keypad or buttons resulting in the inability to utilize the device. Additionally, there is a chance of water ingress or contamination resulting in damage to the internal machinery of the device. Customers are instructed to: 1. Immediately check your internal inventory to locate the product listed on the attached business reply form. 2. Please remove and quarantine the affected product from use. 3. To confirm receipt of this Medical Device Correction Notice and understanding of the provided information, please email the enclosed Business Reply Form (BRF) by January 31st, 2025 date, to RSRecall@Stryker.com. 4. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return and replacement of your product(s). 5. Maintain awareness of this communication internally until all required actions have been completed within your facility. Please ensure this letter is kept with the affected device until replacement. 6. Inform Stryker if any of the subject devices have been distributed to other organizations by contacting Stryker at: a. Device Tracking: rsdevicetracking@stryker.com b. Website: https://www.strykeremergencycare.com/service-support-overview/device-registration/ For question or concerns - contact Customer Service +1 800 442 1142 or by email RSRecall@Stryker.com.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1232-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology