Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1233-2026

EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705 Recall

EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705

FDA recall Class II

Ongoing

FDA lists this Class II recall for EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1233-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 5, 2026
FDA report date
Feb 11, 2026
Quantity in commerce
19
Distribution
See key details
FDA cause
Labeling Change Control

What is being recalled?

EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705

Affected model & UDI checker

Search the exact identifiers published for Z-1233-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
342-14-705EMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705Check source noticeNot specified
Not specifiedEMPOWR 3D Knee Tibial Insert 5R 14MM, VE, REF: 342-14-705Check source notice00888912167529
FDA code information

UDI-DI: 00888912167529. Lot: 139T1075A. Expiration: 03-12-2026

View the original FDA recall notice

Why was this product recalled?

942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.

What actions did Encore Medical, LP take?

On 1/5/2026, recall notices were emailed to customers who were asked to do the following: 1) This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 2) Contact the firm to arrange for affected device return. 3. Complete and return the acknowledgement and response form via email to Lesli.Helmick@enovis.com Customers with questions can email the firm at productsafety@enovis.com

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology