Plato 17 Microcatheter REF PL17-160-090 To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vasculature during interventional or diagnostic procedures. Recall
Plato 17 Microcatheter REF PL17-160-090 To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vasculature during interventional or diagnostic procedures.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Plato 17 Microcatheter REF PL17-160-090 To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vasculature during interventional or diagnostic procedures.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1235-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 16, 2025
- FDA report date
- Mar 5, 2025
- Quantity in commerce
- 1 unit
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Plato 17 Microcatheter REF PL17-160-090 To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vasculature during interventional or diagnostic procedures.
Affected model & UDI checker
Search the exact identifiers published for Z-1235-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| PL17-160-090 | Plato 17 Microcatheter REF PL17-160-090 To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vasculature during interventional or diagnostic procedures. | Check source notice | Not specified |
| Not specified | Plato 17 Microcatheter REF PL17-160-090 To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vasculature during interventional or diagnostic procedures. | Check source notice | 00818075010479 |
FDA code information
Catalog Number: PL17-160-090 UDI-DI code: 00818075010479 Lot Number: 031300
Why was this product recalled?
Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Scientia Vascular, Inc. take?
On 01/02/2025, the firm email an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that through internal evaluation of finished goods of the Plato 17 Microcatheter, one lot of pouches containing the 90 Tip Plato 17 Microcatheter was identified with pouches that had a gap in the sterile barrier seal. Customer are instructed to: In situations where a device has already been used in a patient procedure, closely monitor patients for signs and symptoms of post-surgical infection. Customer Service will be available to provide additional information and provide replacement orders at customercare@scientiavascular.com or (888)385-9016; Monday through Friday, 8:00 AM to 4:30 PM, Mountain Time.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1235-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026