CellMek SPS Sample Preparation System, REF: C44603, with software Recall

CellMek SPS Sample Preparation System, REF: C44603, with software

FDA recall Class II

Ongoing

FDA lists this Class II recall for CellMek SPS Sample Preparation System, REF: C44603, with software. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1235-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 23, 2025
FDA report date
Feb 11, 2026
Quantity in commerce
50
Distribution
See key details
FDA cause
Software change control

What is being recalled?

CellMek SPS Sample Preparation System, REF: C44603, with software

Affected model & UDI checker

Search the exact identifiers published for Z-1235-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
C44603CellMek SPS Sample Preparation System, REF: C44603, with softwareCheck source noticeNot specified
Not specifiedCellMek SPS Sample Preparation System, REF: C44603, with softwareCheck source notice15099590750312
FDA code information

UDI-DI: 15099590750312. Software Version 2.3.90.0

View the original FDA recall notice

Why was this product recalled?

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

What actions did BECKMAN COULTER, INC. take?

On 12/23/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) To mitigate the risk of contact between the dispense probe and the tube bottom, utilize CellMek SPS output tubes distributed by Beckman Coulter. 2) For absolute count assays, use the CellMek SPS on-board counting bead feature. Do not add counting beads to samples manually after preparation with CellMek SPS. 3) Share the content of this letter with your laboratory and/or medical director regarding the need to review previous patient test results. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. 4) Firm recommends posting this letter on or near the affected instruments. 5) Complete and return the response form. Firm will perform an on-site service modification to update your instrument software to correct this issue. If you have any questions regarding this notice, please contact the firm's Customer Support Center: http://www.beckman.com, (800) 369-0333

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology