Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1247-2025

Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Recall

Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.01

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Leica Biosystems Cryostats Product Name: CM1950. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1247-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 30, 2025
FDA report date
Mar 5, 2025
Quantity in commerce
2
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.01

Affected model & UDI checker

Search the exact identifiers published for Z-1247-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
1491950C4USBrand Name: Leica Biosystems Cryostats Product Name: CM1950Check source noticeNot specified
Not specifiedBrand Name: Leica Biosystems Cryostats Product Name: CM1950Check source notice04049188139030
FDA code information

Model No: 1491950C4US UDI-DI: (01)04049188139030 Serial Numbers 12835, 12854

View the original FDA recall notice

Why was this product recalled?

Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did LEICA BIOSYSTEMS NUSSLOCH GMBH take?

On January 30, 2025, the firm notified customers through Email of the recall. Customers were provided with immediate actions to take to mitigate the risk of damaging the tissue sample. Among the firm's recommendations was to have a backup instrument if available. The firm's Service organization will contact customers in the coming weeks to organize the upgrade of affected instruments.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology