Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1248-2026

25GA Subretinal Injection Cannula VS0220.25 Recall

25GA Subretinal Injection Cannula VS0220.25

FDA recall Class II

Ongoing

FDA lists this Class II recall for 25GA Subretinal Injection Cannula VS0220.25. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1248-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 16, 2025
FDA report date
Feb 11, 2026
Quantity in commerce
14,789 (8651 US; 6138 OUS)
Distribution
See key details
FDA cause
Employee error

What is being recalled?

25GA Subretinal Injection Cannula VS0220.25

Affected model & UDI checker

Search the exact identifiers published for Z-1248-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specified25GA Subretinal Injection Cannula VS0220.25Check source notice810123480418
Not specified25GA Subretinal Injection Cannula VS0220.25Check source notice810123480692
FDA code information

Pouch UDI 810123480418 Box UDI 810123480692 Lot 2410043

View the original FDA recall notice

Why was this product recalled?

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Vortex Surgical Inc. take?

On December 16, 2025, the firm began notifying customers via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were instructed to discontinue use of affected product and to return product for Vortex Surgical to evaluate the seals; use UPS account 23623Y to ship unused products to Vortex Surgical Inc for replacement or refund at: Attn: Andrew Ritts Recall Tyvek Seal, Vortex Surgical Inc., 4 Research Park Dr., Suite 124, St. Charles, MO 63304; complete and return the Acknowledgement and Receipt Form via mail or email to: ARitts@VortexSurgical.com. Actions have been implemented to improve training of individuals associated with sealing and inspecting seals to prevent reoccurrence. The sealer associated with the issue is being updated to include a guard to prevent reoccurrence. If you have any questions, contact the Director of Regulatory Affairs at 635-778-4350 or email to: ARitts@VortexSurgical.com from 8am to 5pm.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology