CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056) Recall
CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)
FDA recall Class II
Ongoing
FDA lists this Class II recall for CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056). Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1269-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 9, 2025
- FDA report date
- Mar 12, 2025
- Quantity in commerce
- 30000 ea/ 300 cases
- Distribution
- See key details
- FDA cause
- Process design
What is being recalled?
CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056)
Affected model & UDI checker
Search the exact identifiers published for Z-1269-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | CircuitGuard Bacterial/Viral Filter with Elbow (Model/Catalogue number: SK200FPE / 7056) | Check source notice | 00810050853682 |
FDA code information
UDI (01)00810050853682(10)C2N24002(11)240828(17)290827 Lot C2N24002
Why was this product recalled?
The machine side connector of the filter HME may be occluded by plastic which prevents proper airflow. The issue may lead to ventilation bag malfunction in which the user is unable to inflate/deflate properly during manual ventilation. Potential risks to health include oxygen deprivation, leading to respiratory distress and medical complications.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did Kung Shin Plastics Co. Ltd. take?
On January 9, 2025, the firm sent a recall notice to their sole distributor in the US, followed by a more detailed notice on January 10. Customers were instructed to locate and immediately discontinue use of affected products. Affected product should be quarantined and returned. If products of affected batches have been sold, the distributor should immediately notify the buyers to stop using them, quarantine them, and destroy them.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1269-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026