Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1284-2025

smiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mm O.D. (7.8Fr) x 1.6mm I.D., REF 21-4172-24 Recall

smiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mm O.D. (7.8Fr) x 1.6mm I.D., REF 21-4172-24

FDA recall Class I

Ongoing

FDA lists this Class I recall for smiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mm O.D. (7.8Fr) x 1.6mm I.D., REF 21-4172-24. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1284-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Feb 13, 2025
FDA report date
Mar 26, 2025
Quantity in commerce
1 unit
Distribution
See key details
FDA cause
Process control

What is being recalled?

smiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mm O.D. (7.8Fr) x 1.6mm I.D., REF 21-4172-24

Affected model & UDI checker

Search the exact identifiers published for Z-1284-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
21-4172-24smiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mm O.D. (7.8Fr) x 1.6mm I.D., REF 21-4172-24Check source noticeNot specified
Not specifiedsmiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mm O.D. (7.8Fr) x 1.6mm I.D., REF 21-4172-24Check source notice10610586012723
FDA code information

UDI/DI 10610586012723, Lot Numbers: 4235509

View the original FDA recall notice

Why was this product recalled?

The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect.

What actions did Smiths Medical ASD, Inc. take?

Smiths medical issued an URGENT MEDICAL DEVICE RECALL to its consignees on 02/13/2025 via letter delivered by traceable means. The notice explained the issue, potential risk, and requested the following: Required Actions for Customers 1. Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2. Please return all affected product. 3. Inform all potential users of the product in your organization of this notification and complete and return the attached response form to smithsmedical8171@sedgwick.com within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product and/or has already been used. 4. If you have distributed the product further, immediately notify your accounts that received the affected product and ask them to contact Sedgwick at 1-888-345-2656 (MF, 8am-5pm ET) to obtain a response form.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology