smiths medical ProPort Low Profile Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm (5.8Fr) O.D. x 1.0mm., 6Fr Introducer, REF 21-4183-24 Recall
smiths medical ProPort Low Profile Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm (5.8Fr) O.D. x 1.0mm., 6Fr Introducer, REF 21-4183-24
FDA recall Class I
Ongoing
FDA lists this Class I recall for smiths medical ProPort Low Profile Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm (5.8Fr) O.D. x 1.0mm., 6Fr Introducer, REF 21-4183-24. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1287-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Feb 13, 2025
- FDA report date
- Mar 26, 2025
- Quantity in commerce
- 8515 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
smiths medical ProPort Low Profile Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm (5.8Fr) O.D. x 1.0mm., 6Fr Introducer, REF 21-4183-24
Affected model & UDI checker
Search the exact identifiers published for Z-1287-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 21-4183-24 | smiths medical ProPort Low Profile Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm (5.8Fr) O.D. x 1.0mm., 6Fr Introducer, REF 21-4183-24 | Check source notice | Not specified |
| Not specified | smiths medical ProPort Low Profile Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm (5.8Fr) O.D. x 1.0mm., 6Fr Introducer, REF 21-4183-24 | Check source notice | 10610586012778 |
FDA code information
UDI/DI 10610586012778, Lot Numbers: 3916028, 3969277, 4006477, 4022602, 4036895, 4057818, 4060894, 4096338, 4108453, 4114117, 4146468, 4153874, 4163556, 4196775, 4221727, 4235565, 4235567, 4235601, 4248694, 4256927, 4256928, 4256964, 4276228, 4294894, 4294895, 4364693, 4366695, 4366696, 4378523, 4382349, 4392370, 4395513, 4405010, 4420761, 4426418, 4434230, 4453606, 4468579, 4468580, 6005957, 6005958, 6026650, 6026651, 6037122, 6037124, 6051325.
Why was this product recalled?
The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect.
What actions did Smiths Medical ASD, Inc. take?
Smiths medical issued an URGENT MEDICAL DEVICE RECALL to its consignees on 02/13/2025 via letter delivered by traceable means. The notice explained the issue, potential risk, and requested the following: Required Actions for Customers 1. Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2. Please return all affected product. 3. Inform all potential users of the product in your organization of this notification and complete and return the attached response form to smithsmedical8171@sedgwick.com within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product and/or has already been used. 4. If you have distributed the product further, immediately notify your accounts that received the affected product and ask them to contact Sedgwick at 1-888-345-2656 (MF, 8am-5pm ET) to obtain a response form.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1287-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology