Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1300-2026

Bubble Sensor for 3/8" x 3/32" tubing; Recall

Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

FDA recall Class II

Ongoing

FDA lists this Class II recall for Bubble Sensor for 3/8" x 3/32" tubing;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1300-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 9, 2026
FDA report date
Feb 11, 2026
Quantity in commerce
3050 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

Affected model & UDI checker

Search the exact identifiers published for Z-1300-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
701055720Bubble Sensor for 3/8" x 3/32" tubing;Check source noticeNot specified
Not specifiedBubble Sensor for 3/8" x 3/32" tubing;Check source notice04037691816432
FDA code information

Model Number: 701055720; UDI-DI: 04037691816432; Serial Numbers: All units impacted;

View the original FDA recall notice

Why was this product recalled?

Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.

What actions did Maquet Cardiopulmonary Gmbh take?

On or around January 9, 2026, "URGENT MEDICAL DEVICE RECALL" letters were sent to customers. Customers were instructed to take the following actions: - Please examine your inventory immediately to determine if you have any affected product in your inventory and utilize the recommended mitigations listed above until replacement sensors are available. - Complete and return the enclosed response form Getinge by e-mailing a copy to recallresponses.qrc@getinge.com. - Contact Customer Service via email at ACTSalesSupport.US@Getinge.com or 888-943-8872, ext 28849 / (888 9GETUSA), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. (Eastern Time Zone) to order a replacement sensor that will be shipped once available. Affected sensors will be returned to Getinge at time of replacement. - Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. If you have any questions, contact Customer Service via email at ACTSalesSupport.US@Getinge.com or 888-943-8872, ext 28849 / (888 9GETUSA), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. (Eastern Time Zone).

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology