Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Recall
Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number: CL0073 Software Version: Not Applicable Product Description: Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio is an in vitro diagnostic (IVD) test that combines the test results of the Lumipulse G pTau 217 Plasma assay and the Lumipulse G ¿-Amyloid 1-42-N Plasma assay from the same patient specimen (K2EDTA plasma sample) into a numerical ratio. The Lumipulse G pTau 217/¿-Amyloid 1-42 Plasma Ratio Instructions For Use (IFU), CL0072, contains instructions on how to manually calculate the ratio and is supplied with each individual Lumipulse G assay. Lumipulse G pTau217/¿-Amyloid 1-42 Plasma Ratio, Item# CL0073 is comprised of the following kits, which are imported: Kit Description, Item#: Imported from Fujirebio Europe (FRE): Lumipulse G pTau 217 Plasma Immunoreaction Cartridges 81550 Lumipulse G pTau 217 Plasma Calibrators set 81557 Lumipulse pTau 217 Plasma Controls 81558 Lumipulse ¿-Amyloid Plasma Controls 81559 Imported from Fujirebio Japan (FRI): Lumipulse G ¿-Amyloid 1-42-N Plasma Immunoreaction Cartridges 235447 Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators set 235454 Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1301-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 11, 2025
- FDA report date
- Feb 11, 2026
- Quantity in commerce
- 23 units
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio Model/Catalog Number: CL0073 Software Version: Not Applicable Product Description: Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio is an in vitro diagnostic (IVD) test that combines the test results of the Lumipulse G pTau 217 Plasma assay and the Lumipulse G -Amyloid 1-42-N Plasma assay from the same patient specimen (K2EDTA plasma sample) into a numerical ratio. The Lumipulse G pTau 217/-Amyloid 1-42 Plasma Ratio Instructions For Use (IFU), CL0072, contains instructions on how to manually calculate the ratio and is supplied with each individual Lumipulse G assay. Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio, Item# CL0073 is comprised of the following kits, which are imported: Kit Description, Item#: Imported from Fujirebio Europe (FRE): Lumipulse G pTau 217 Plasma Immunoreaction Cartridges 81550 Lumipulse G pTau 217 Plasma Calibrators set 81557 Lumipulse pTau 217 Plasma Controls 81558 Lumipulse -Amyloid Plasma Controls 81559 Imported from Fujirebio Japan (FRI): Lumipulse G -Amyloid 1-42-N Plasma Immunoreaction Cartridges 235447 Lumipulse G -Amyloid 1-42-N Plasma Calibrators set 235454 Component: No
Affected model & UDI checker
Search the exact identifiers published for Z-1301-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| CL0073 | Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio | Check source notice | Not specified |
| Not specified | Product Name: Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio | Check source notice | 00869164000086 |
FDA code information
Model/Catalog Number: CL0073; UDI-DI: 00869164000086; Distribution for Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio ("Ratio") is related to the "Ratio" Instructions for Use (IFU). Therefore, each consignee has 1 "Ratio" in which they purchase the component kits for.
Why was this product recalled?
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
What actions did Fujirebio Diagnostics, Inc. take?
On or about December 11, 2025 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Immediately discontinue using the associated Lots listed above for the Lumipulse G pTau217/-Amyloid 1-42 Plasma Ratio. - Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. - Notify your ordering physician or health care provider as appropriate. - The test results must be interpreted in conjunction with other diagnostic tools and clinical information. - Please retain this letter with your laboratory records and forward this leer to those who may have received this product.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1301-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026