Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1314-2025

COR Disposable Kit, 8 mm. Cartilage Transplant System. Recall

COR Disposable Kit, 8 mm. Cartilage Transplant System.

FDA recall Class II

Ongoing

FDA lists this Class II recall for COR Disposable Kit, 8 mm. Cartilage Transplant System.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1314-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 24, 2025
FDA report date
Mar 19, 2025
Quantity in commerce
US: 131 units; OUS: 312 units
Distribution
See key details
FDA cause
Employee error

What is being recalled?

COR Disposable Kit, 8 mm. Cartilage Transplant System.

Affected model & UDI checker

Search the exact identifiers published for Z-1314-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
252109COR Disposable Kit, 8 mm. Cartilage Transplant System.Check source noticeNot specified
Not specifiedCOR Disposable Kit, 8 mm. Cartilage Transplant System.Check source notice10886705012374
FDA code information

Model No. 252109 & 252112; GTIN: 10886705012374 & 10886705012404; Lot No. 23C11, 23C12, 23E04, 23F02, 23R03, 24A02, 24B01, 24C17, 24P07, 23C13, 23D03, 23D14, 23F01, 24C15, 24K07 & 24N10.

View the original FDA recall notice

Why was this product recalled?

Device is missing the pin in the graft loader component.

What actions did DePuy Mitek, Inc., a Johnson & Johnson Co. take?

Consignees were sent an URGENT MEDICAL DEVICE RECALL (REMOVAL) notification dated 2/20/25. The notification instructs consignees to examine their inventory and quarantine any affected devices. Recalled product is not to be used. Once affected devices have been quarantined, consignees are to contact their DePuy Synthes Sales Consultant to coordinate their return. Consignees are to complete and return the provided Business Response Form to the provided email address within three business days of receipt. The recall notification should be shared with those in your organization that need to be informed and forwarded to facilities to which product was further distributed. The recall notification should also be posted in a visible area for awareness. Consignees with any questions are to contact their DePuy Synthes Sales Consultant.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.