Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1329-2025

Z1 Femoral Hip System, Product Number 611777612 Recall

Z1 Femoral Hip System, Product Number 611777612

FDA recall Class II

Ongoing

FDA lists this Class II recall for Z1 Femoral Hip System, Product Number 611777612. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1329-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 12, 2025
FDA report date
Mar 19, 2025
Quantity in commerce
35 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Z1 Femoral Hip System, Product Number 611777612

Affected model & UDI checker

Search the exact identifiers published for Z-1329-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedZ1 Femoral Hip System, Product Number 611777612Check source notice00887868577888
FDA code information

UDI-DI: 00887868577888; Lot Numbers: 102646 102777 102778 104095 104324 105277

View the original FDA recall notice

Why was this product recalled?

There is a potential design issue with the Z1 Offset Stem Inserter leading to fracturing of the distal tip.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Zimmer, Inc. take?

An URGENT MEDICAL DEVICE RECALL notification dated 2/12/25 was sent to customers. Our records indicate that you may have received one or more of the affected products. The affected units were distributed between September 2024 and January 2025. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. For cases where the Z1 Offset Stem Inserter was used, increased clinical or radiographic monitoring may be required if the device fractured during use. If the Z1 Offset Stem Inserter was not used or if there was no evidence of fracture during use, there are no additional specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA classification recorded

  3. FDA enforcement report date

  4. FDA notice posted

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology