Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1334-2025

MAJOR DIEP, Model Nos. CCMJ73G-01 CCMJ73H Recall

MAJOR DIEP, Model Nos. CCMJ73G-01 CCMJ73H

FDA recall Class II

Ongoing

FDA lists this Class II recall for MAJOR DIEP, Model Nos. CCMJ73G-01 CCMJ73H. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1334-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 6, 2025
FDA report date
Mar 26, 2025
Quantity in commerce
17 kits
Distribution
See key details
FDA cause
Process control

What is being recalled?

MAJOR DIEP, Model Nos. CCMJ73G-01 CCMJ73H

Affected model & UDI checker

Search the exact identifiers published for Z-1334-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMAJOR DIEP, Model Nos. CCMJ73G-01 CCMJ73HCheck source notice00191072222824
Not specifiedMAJOR DIEP, Model Nos. CCMJ73G-01 CCMJ73HCheck source notice00191072223890
FDA code information

UDI-DI 00191072222824 00191072223890 Lots 879241 70-052915

View the original FDA recall notice

Why was this product recalled?

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

What actions did AMERICAN CONTRACT SYSTEMS, INC. take?

The affected distributor was notified via email on 2/16/25, and an updated letter was provided on 2/18/25. The distributor was instructed to notify their end users. Customers were provided with a list of impacted components which are included in the affected trays. Customers were provided with a labeling template and instructed to add warning labels to affected procedure trays to ensure the affected components are discarded.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology