Halyard TRANSPORT BAG KIT. Kit Code: LIFE0080-01. Recall
Halyard TRANSPORT BAG KIT. Kit Code: LIFE0080-01.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Halyard TRANSPORT BAG KIT. Kit Code: LIFE0080-01.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1338-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 2, 2026
- FDA report date
- Feb 18, 2026
- Quantity in commerce
- 1,500 units
- Distribution
- See key details
- FDA cause
- Equipment maintenance
What is being recalled?
Halyard TRANSPORT BAG KIT. Kit Code: LIFE0080-01.
Affected model & UDI checker
Search the exact identifiers published for Z-1338-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Halyard TRANSPORT BAG KIT. Kit Code: LIFE0080-01. | Check source notice | 10809160314500 |
FDA code information
Kit Code: LIFE0080-01. UDI-DI: 10809160314500. Lot Number: 1650056. Expiration Date: 07/24/2028
Why was this product recalled?
Tyvek bag seal issue which may compromise sterility of the kit.
What actions did AVID Medical, Inc. take?
AVID Medical notified consignees on about 01/02/2026 via emailed URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter. Consignees were instructed to segregate and quarantine all affected lots, discard all affected lots on hand, and complete and return the provided Recall Response form. Distributors were instructed to notify all customers if further distributed.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1338-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology