Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Recall
Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;
FDA recall Class II
Ongoing
FDA lists this Class II recall for Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1341-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 13, 2026
- FDA report date
- Feb 18, 2026
- Quantity in commerce
- 1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)
- Distribution
- See key details
- FDA cause
- Nonconforming Material/Component
What is being recalled?
Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;
Affected model & UDI checker
Search the exact identifiers published for Z-1341-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 880052 | Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; | Check source notice | Not specified |
| 880062 | Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; | Check source notice | Not specified |
| 881152 | Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; | Check source notice | Not specified |
| Not specified | Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; | Check source notice | 07045432088519 |
| Not specified | Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; | Check source notice | 07045432088533 |
| Not specified | Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; | Check source notice | 07045432067132 |
FDA code information
1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;
Why was this product recalled?
units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.
What actions did Laerdal Medical Corporation take?
On January 13, 2026, Medical Device Recall letters were sent to customers. Actions to be taken: 1. Please review this notice carefully and provide acknowledgment that you have received and understood the contents. If you received this notice from a distributor or other third party, please send your acknowledgment directly to Laerdal using the contact information provided below. You must provide the case number in the footer of this notice in your response. 2. Ensure that this notice is shared with all relevant personnel involved in the storage, maintenance, charging, and use of the LCSU 4 RTCA device. 3. Maintain this information for your records. As part of Laerdal s commitment to customer service, RTCA/DO-160G Section 21 Category M compliant replacement units are being offered, free of charge. To receive a replacement, you must reply within 2 weeks of receipt, with the following information: 1. Confirm acknowledgment of this notice as mentioned above. 2. Confirm serial numbers of your affected units, shipping address, and contact information in your reply. Once validated, Laerdal Medical will arrange for your affected unit(s) to be replaced at no cost. Important: Replacement units (881152) are shipped without a carry bag, battery, external power supply, wire stand, canister, or tubing. Please retain these accessories from your existing unit for continued use. Customers are required to confirm responsibility for disposing of the replaced LCSU 4 RTCA Unit(s) in accordance with applicable environmental regulations. Proof of disposal must be retained and made available upon request. By accepting the replacement, the customer releases Laerdal from any future claims or liabilities related to the old unit. Please maintain awareness of this notice and the related actions for your records. ***Update 03/24/2026***Additional recall letters were sent on or around 2/25/2026 to additional consignees identified by the recalling firm.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1341-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026