da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45 Recall

da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

FDA recall Class II

Ongoing

FDA lists this Class II recall for da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1355-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 30, 2025
FDA report date
Feb 18, 2026
Quantity in commerce
47
Distribution
See key details
FDA cause
Process change control

What is being recalled?

da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

Affected model & UDI checker

Search the exact identifiers published for Z-1355-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedda Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45Check source notice00886874119747
FDA code information

UDI: 00886874119747, Serial: SQ0326, SQ0336, SQ0341, SQ0345, SQ0346, SQ0356, SQ0367, SQ0368, SQ0375, SQ0377, SQ0379, SQ0380, SQ0386, SQ0387, SQ0389, SQ0394, SQ0395, SQ0398, SQ0401, SQ0404, SQ0412, SQ0413, SQ0415, SQ0419, SQ0420, SQ0422, SQ0425, SQ0427, SQ0428, SQ0432, SQ0433, SQ0436, SQ0443, SQ0447, SQ0448, SQ0451, SQ0454, SQ0457, SQ0458, SQ0462, SQ0464, SQ0469, SQ0470, SQ0662, SQ0727, SQ0740, SQ0761

View the original FDA recall notice

Why was this product recalled?

Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Intuitive Surgical, Inc. take?

On 12/30/2025, correction notices were mailed and emailed to customers who were informed the following: If single-viewer failure occurs during a procedure, when clinically safe to do so, restart the system to restore 3D visualization in the viewer. If dual-viewer failure occurs during a procedure, be aware that instrument motion will not be automatically halted by the system. Instrument motion can be halted by removing the surgeon s head from the viewer. The surgeon maintains the ability to control the instruments while their head is in the viewer. To ensure instruments do not cause unintended tissue interaction, the bedside user should maneuver installed instruments into a safe space in the surgical field, using an external monitor for visualization. Once the instruments are safely positioned, restart the system to restore visualization. At any point during the sequence, the surgeon may press E Stop to put the system in a safe state (arms locked, no instrument motion). If any viewer display issue is detected, the user should avoid starting any additional procedures with the affected system until firm has repaired the system. Forward this notification to all personnel at your site who perform and support these procedures. Complete and return the acknowledgement form via email to Recalls@intusurg.com Retain a copy of this letter and the acknowledgement form for your files. Firm's Field Engineers will contact affected customers to schedule a site visit to upgrade the Viewer Display firmware on affected systems. If you need further information or support concerning this Urgent Medical Device Correction, please contact your Clinical Sales Representative or contact the firm's Customer Service at: North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: customerservice@intusurg.com.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology