FDA RECALL RECORD · Z-1389-2025

Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CE Recall

Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CE

FDA recall Class II

Ongoing

FDA lists this Class II recall for Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CE. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1389-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 10, 2025
FDA report date
Mar 26, 2025
Quantity in commerce
30 vials
Distribution
See key details
FDA cause
Process change control

What is being recalled?

Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CE

Affected model & UDI checker

Search the exact identifiers published for Z-1389-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
C50110Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CECheck source noticeNot specified
ID1001Candida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CECheck source noticeNot specified
Not specifiedCandida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CECheck source notice00816387020223
FDA code information

Lot #: F4171005/UDI: 00816387020223

View the original FDA recall notice

Why was this product recalled?

Candida Immunodiffusion Antigen may have reduced sensitivity which could result in false negative results.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Immuno-Mycologics, Inc take?

On February 10, 2025, Immy issued a "Urgent Medical Device Recall" Notification to affected consignees. Immy asked consignees to take the following actions: 1. Immediately identify, count, and segregate any affected vials you have in your inventory to prevent them from being used or shipped to your customers. As soon as possible, discard your affected inventory. 2. Complete the attached Acknowledgement and Receipt Form (pages 3 and 4 below) even if you do not have any affected stock remaining in your possession. Note: The form is a fillable PDF. You can save it to your computer, fill out electronically and attach to an email. Return the completed form to IMMY using one of the methods below: o Email: customerservice@immy.com; Mail to: Attn: Hunter Conover IMMY, Inc., 2701 Corporate Centre Dr., Norman, OK USA 73069 3. Ensure relevant staff members are informed of this recall, including relevant clinicians. Clinicians should review all negative results. 4. If you have supplied any potentially affected product to another organization, please advise that organization of this recall and send them this notification. Please contact us so we can follow up with them. 5. In case product is in transit, display this letter in a prominent place for one month.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology